Comparison of fluorescence parameters between three generations of QLF devices for detecting enamel caries in vitro
Seok-Woo Park1, Sang-Kyeom Kim1, Hyung-Suk Lee1
1Department of Preventive Dentistry & Public Oral Health, BK21 PLUS project, Yonsei University College of Dentistry, Seoul, Republic of Korea.
Background:
This study compared two fluorescence parameters (fluorescence loss [ΔF] and red fluorescence gain [ΔR]) among three generations of quantitative light-induced fluorescence (QLF) systems with the aim of determining the validities of these parameters in the three devices for differentiating the severity of enamel caries.
Methods:
Forty-one extracted human premolars and molars with suspected enamel caries were selected. Fluorescence images of all teeth were obtained using first-, second-, and third-generation QLF systems (Inspektor Pro, QLF-D, and Qraycam, respectively). Fluorescence parameters were then calculated using proprietary software. All of the specimens were also categorized histologically using polarized-light microscopy (PLM) based on histological levels related to the lesion depth into sound enamel (S), caries limited to the outer half of the enamel (E1), and caries involving the inner half of the enamel (E2). The Mann-Whitney test with Bonferroni correction was used to compare fluorescence parameters among the three generations of systems. The sensitivity, specificity, and area under the receiver operating characteristics curve (AUC) at two thresholds (S/E1 for detecting enamel caries lesions and E1/E2 for differentiating the caries severity) were calculated for evaluating the validities of the fluorescence parameters obtained using all three generations of QLF devices.
Results:
ΔF did not differ significantly between the devices at any histological level. In addition, ΔF showed large AUCs at the thresholds of S/E1 and E1/E2 (0.97-0.98 and 0.89-0.90, respectively). On the other hand, ΔR was significantly higher for the third-generation device than for the first- and second-generation devices for E2 lesions (P < 0.001). At the S/E1 threshold, ΔR values of the first- and third-generation devices showed larger AUCs (0.96-0.97) compared with that of the second-generation device (0.91), whereas at the E1/E2 threshold the AUC was the largest for the third-generation device (0.87).
Conclusions:
The ΔF fluorescence parameter did not differ between the three generations of QLF devices, and showed high validity values. In terms of ΔR, the devices of all generations also showed good diagnostic performance for quantifying and detecting enamel caries lesions, but the third-generation QLF system produced superior results.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Smooth Endoplasmic Reticulum
The ER provides optimal conditions for synthesizing steroid hormones and lipids, such as phospholipids and triglycerides. Traditionally, lipid metabolism was considered to be a smooth ER function. However, there is no direct evidence to prove that rough ER is completely excluded from lipid...
Functions of Smooth Muscles
Function of visceral smooth muscles
Visceral smooth muscle is found in the walls of all hollow organs, except the heart, and is a key player in the involuntary movements that drive the functioning of these internal organs. This tissue is arranged in...
Smooth Muscle Contraction
The onset of contraction is triggered by an increase in calcium ions within the sarcoplasm, similar to the process in striated muscle. However, smooth muscles have a relatively smaller reservoir of the sarcoplasmic...
Wave Parameters
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...