Quarter-millimeter spectral coronary stent imaging with photon-counting CT: Initial experience
Rolf Symons1, Yves De Bruecker2, John Roosen3
1Radiology and Imaging Sciences - National Institutes of Health Clinical Center, Bethesda, MD, USA; Department of Imaging & Pathology, University Hospitals Leuven, Leuven, Belgium.
Purpose:
To evaluate the performance and clinical feasibility of 0.25 mm resolution mode of a dual-energy photon-counting detector (PCD) computed tomography (CT) system for coronary stent imaging and to compare the results to state-of-the-art dual-energy energy-integrating detector (EID) CT.
Materials And Methods:
Coronary stents with different diameters (2.0-4.0 mm) were examined inside a coronary artery phantom consisting of plastic tubes filled with iodine-based and gadolinium-based contrast material diluted to approximate clinical concentrations (n = 18). EID images were acquired using 2nd and 3rd generation dual-source CT systems (SOMATOM Flash and SOMATOM Force, Siemens Healthcare) at 0.60 mm (defined as standard-resolution (SR)) isotropic voxel size. Radiation-dose matched PCD images were acquired using a human prototype PCD system (Siemens Healthcare) at 0.50 mm (SR) and 0.25 mm (HR) imaging modes. Images were reconstructed using optimized convolution kernels.
Results:
Dual-energy HR PCD images significantly better stent lumen visualization (median: 69.5%, IQR: 61.2-78.9%) over dual-energy EID, and standard-resolution PCD images (median: 53.2-57.4%, all P < 0.01). HR PCD acquisitions reconstructed at SR image voxel size showed 25.3% lower image noise compared to SR PCD acquisitions (P < 0.001). High-resolution iodine and gadolinium maps, as well as virtual monoenergetic images, were calculated from the PCD data and enabled estimation of contrast agent concentration in the lumen without interference from the coronary stent.
Conclusion:
HR spectral PCD imaging significantly improves coronary stent lumen visibility over dual-energy EID. When the PCD-HR data was reconstructed into standard voxel sizes (0.5 mm isotropic) the image noise decreased by 25% compared to SR acquisition of PCD. Both dual-energy systems were consistent in estimating contrast agent concentrations.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...