A Pilot Study on Parameter Setting of VisiTag™ Module during Pulmonary Vein Isolation
Yu-Chuan Wang1, Bo Huang1, Kang Li2
1Department of Geriatrics, Peking University First Hospital, Beijing, China.
Optimal predefined criteria for VisiTag™ module settings during pulmonary vein isolation (PVI) were identified. A 3mm distance limit and specific force-over-time settings minimize gaps for effective atrial fibrillation treatment.
Area of Science:
- Electrophysiology
- Cardiac Ablation Technology
Background:
- Pulmonary vein isolation (PVI) is a key treatment for atrial fibrillation (AF).
- The CARTO 3 system with its VisiTag™ module aids in real-time ablation lesion assessment.
- Optimizing predefined criteria (OPC) for the VisiTag™ module is crucial for successful PVI.
Purpose of the Study:
- To determine the optimal predefined criteria (OPC) for the VisiTag™ module filters.
- To minimize conduction gaps (CG) and maximize contiguous encircling ablation lines (CEAL) during PVI.
Main Methods:
- Thirty AF patients undergoing PVI using a Thermocool SmartTouch catheter were enrolled.
- Ablation lesions were automatically tagged based on VisiTag™ module criteria.
- Different parameter settings, including catheter movement limits and force-over-time (FOT), were evaluated.
Main Results:
- Initial settings (3mm distance limit, 20s, FOT off) yielded 60 CEAL but 26 CGs.
- Adjusting FOT to 5g/50% stability resulted in 22 tag gaps (TG), 20 with CGs.
- Reablation at TG sites eliminated most CGs, but aggressive FOT settings (10g/50%) eliminated CEAL.
Conclusions:
- A 3mm distance limit for at least 20s combined with FOT set to a minimal force of 5g and 50% stability time are proposed as OPC.
- These settings appear to balance lesion completeness and minimize gaps during PVI.
Related Concept Videos
Pilot and Numeric Relaying
Varicose Veins II: Diagnostic Studies and Interprofessional Care
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Wave Parameters
Veins of Thorax
The azygos vein, positioned just right of the midline and anterior to the vertebral column, begins at the junction of the right ascending lumbar and subcostal veins, terminating in the superior vena cava. This vein drains blood from the right side of the thoracic wall, thoracic viscera, and posterior abdominal wall.
The...