Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Leaky Scanning02:28

Leaky Scanning

5.7K
During most eukaryotic translation processes, the small 40S ribosome subunit scans an mRNA from its 5' end until it encounters the first start AUG codon. The large 60S ribosomal subunit then joins the smaller one to initiate protein synthesis. The location of the translation initiation is largely determined by the nucleotides near the start codon as there may be multiple translation initiation sites present on the mRNA.  Marilyn Kozak discovered that the sequence RCCAUGG (where R...
5.7K
Reducing Line Loss01:18

Reducing Line Loss

391
In a three-phase circuit, line loss is an indicator of energy dissipated as heat due to the resistance of transmission lines. To address this, incorporating transformers into the system—a step-up transformer at the source and a step-down transformer at the load—is a strategic solution. Two three-phase transformers are introduced to improve this.
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
391
Methods of reducing fever01:22

Methods of reducing fever

1.4K
The signs and symptoms of fever include hot and dry skin, flushed face, thirst, muscle aches, anorexia, headache, tachycardia, tachypnea, and fatigue. Elevated body temperature is reduced using two methods: pharmacological and nonpharmacological. Proper identification and treatment of the root cause of a fever is of utmost importance.
Pharmacological Methods of Reducing Fever:
1.4K
Adaptations that Reduce Water Loss01:57

Adaptations that Reduce Water Loss

28.2K
Though evaporation from plant leaves drives transpiration, it also results in loss of water. Because water is critical for photosynthetic reactions and other cellular processes, evolutionary pressures on plants in different environments have driven the acquisition of adaptations that reduce water loss.
28.2K
Scanning Electron Microscopy01:07

Scanning Electron Microscopy

5.5K
A scanning electron microscope (SEM) is used to study the surface features of a sample by using an electron beam that scans the sample surface in a two-dimensional manner. Typically, areas between ~1 centimeter to 5 micrometers in width can be imaged. SEM can be used to image bacteria, viruses, tissues as well as larger samples like insects. Conventional SEM gives a magnification ranging from 20X to 30,000X and spatial resolution of 50 to 100 nanometers.
Fundamental Principles
Accelerated...
5.5K
Role of Reduced Coenzymes NADH and FADH₂01:29

Role of Reduced Coenzymes NADH and FADH₂

16.6K
The energy released from the breakdown of the chemical bonds within nutrients can be stored either through the reduction of electron carriers or in the bonds of adenosine triphosphate (ATP). In living systems, a small class of compounds functions as mobile electron carriers, molecules that bind to and shuttle high-energy electrons between compounds in pathways. The principal electron carriers that will be considered originate from the B vitamin group and are derivatives of nucleotides; they are...
16.6K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Labilized ligands under ultra-confinement: a solid-state <sup>2</sup>H NMR investigation of Ln(BH<sub>4</sub>)<sub>2</sub>(THF)<sub>2</sub>/HY<sub>30</sub>.

Chemical communications (Cambridge, England)·2026
Same author

Group 4 Metallocenes Supported on Sulfated Zirconium Oxide Catalyze Benzene C-H Borylation.

Journal of the American Chemical Society·2026
Same author

Regioselective Hydrosilylation Catalysis with Supported Well-Defined Pt(0) Complexes: Effects of Surface Anions and Phosphenium Ligands.

Inorganic chemistry·2026
Same author

Probing the Surface Structure of Amine/Chloride-Passivated CdSe Nanocrystals Using Dynamic Nuclear Polarization-Enhanced <sup>15</sup>N, <sup>35</sup>Cl, <sup>77</sup>Se, and <sup>113</sup>Cd Solid-State NMR Spectroscopy.

Journal of the American Chemical Society·2026
Same author

Advances in solid-state NMR for the studies of mesoporous solids: fast magic angle spinning and dynamic nuclear polarization.

Progress in nuclear magnetic resonance spectroscopy·2025
Same author

DFT calculations and theory do not support enantiospecificity in NMR J-coupling constants.

Nature communications·2025

Related Experiment Video

Updated: Feb 1, 2026

Automated Segmentation of Cortical Grey Matter from T1-Weighted MRI Images
06:48

Automated Segmentation of Cortical Grey Matter from T1-Weighted MRI Images

Published on: January 7, 2019

9.5K

Reducing t1 noise through rapid scanning.

Frédéric A Perras1, Marek Pruski2

  • 1US DOE, Ames Laboratory, Ames, IA 50011, USA.

Journal of Magnetic Resonance (San Diego, Calif. : 1997)
|December 5, 2018
PubMed
Summary
This summary is machine-generated.

Optimize 2D NMR sensitivity by adjusting relaxation delays. A shorter delay (0.2T1) significantly boosts signal detection in experiments like 1H{13C} D-HMQC, nearly doubling sensitivity compared to conventional methods.

Keywords:
2D NMRD-HMQCSolid-state NMRSpin-lattice relaxationt(1) noise

More Related Videos

Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo
08:01

Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo

Published on: September 26, 2016

9.8K
Fine-tuning the Size and Minimizing the Noise of Solid-state Nanopores
09:43

Fine-tuning the Size and Minimizing the Noise of Solid-state Nanopores

Published on: October 31, 2013

14.2K

Related Experiment Videos

Last Updated: Feb 1, 2026

Automated Segmentation of Cortical Grey Matter from T1-Weighted MRI Images
06:48

Automated Segmentation of Cortical Grey Matter from T1-Weighted MRI Images

Published on: January 7, 2019

9.5K
Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo
08:01

Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo

Published on: September 26, 2016

9.8K
Fine-tuning the Size and Minimizing the Noise of Solid-state Nanopores
09:43

Fine-tuning the Size and Minimizing the Noise of Solid-state Nanopores

Published on: October 31, 2013

14.2K

Area of Science:

  • Nuclear Magnetic Resonance (NMR) Spectroscopy
  • Solid-State NMR
  • Spectrometer Hardware and Noise Analysis

Background:

  • T1 noise, originating from spectrometer hardware instabilities, is a major limitation for sensitivity in 2D NMR experiments.
  • This noise particularly impacts solid-state NMR methods employing dipolar-recoupling techniques.
  • The linear scaling of T1 noise with signal intensity necessitates a re-evaluation of conventional experimental parameters.

Purpose of the Study:

  • To investigate the relationship between signal intensity and T1 noise in 2D NMR experiments.
  • To determine optimal relaxation delay times for maximizing sensitivity in 2D NMR, challenging conventional 1.3T1 settings.
  • To demonstrate improved sensitivity in specific 2D NMR experiments by employing shorter relaxation delays.

Main Methods:

  • Revisiting the theoretical relationship between signal intensity and T1 noise.
  • Comparing sensitivity gains using conventional versus optimized (shorter) relaxation delays.
  • Acquisition of 1H{13C} Dipolar-mediated Heteronuclear Multiple-Quantum Correlation (D-HMQC) spectra.

Main Results:

  • Conventional relaxation delays (e.g., 1.3T1) are often suboptimal for 2D NMR sensitivity due to T1 noise characteristics.
  • A faster repetition rate, enabled by shorter relaxation delays (e.g., 0.2T1), can significantly enhance time sensitivity.
  • Nearly double the sensitivity was achieved in 1H{13C} D-HMQC spectra using a 0.2T1 relaxation delay.

Conclusions:

  • Optimizing the relaxation delay based on noise fluctuation timescales, rather than solely on T1 relaxation, is crucial for 2D NMR sensitivity.
  • Shorter relaxation delays (e.g., 0.2T1) are recommended for 2D experiments when noise fluctuations are faster than the recycle delay.
  • This strategy offers a practical method to nearly double sensitivity in sensitive solid-state NMR experiments like D-HMQC.