Comparison of the novel ARCHITECT procalcitonin assay with established procalcitonin assay systems
1Medical and Scientific Affairs, Abbott Laboratories, Abbott Park, IL, USA.
Aims:
This study assessed the performance of a new fully automated immunoassay, ARCHITECT B.R.A.H.M.S procalcitonin (PCT), comparing the results with other commercial assays on routine clinical specimens.
Methods:
At nine sites from eight countries, precision analysis was carried out on controls by ANOVA. Threshold and linearity were verified according to standard procedures. Comparison of ARCHITECT B.R.A.H.M.S PCT with the Cobas®, LIAISON®, VIDAS® and Kryptor® PCT assays was evaluated using Passing-Bablok and Deming regression analyses.
Results:
The within-laboratory standard deviation and %CV across all sites ranged from 0.005 to 0.008 and 2.7 to 4.1; 0.040 to 0.212 and 2.1 to 11.7; 1.628 to 4.191 and 2.5-6.3 for the three control levels, respectively. The mean slope (linearity analysis) across all sites ranged from 0.85 to 1.03, with a mean y-intercept ranging from -6.15 to + 1.71 and a correlation coefficient ranging from 0.94 to 1.00. The LoB, LoD, and LoQ claims were verified. Deming regression analysis of 1116 plasma or serum samples with PCT results detected across a dynamic assay range of 0.02-100 μg/l using the ARCHITECT B.R.A.H.M.S PCT assay yielded results of r = 0.989 vs. Roche Cobas®, r = 0.986 vs Kryptor® B.R.A.H.M.S, r = 0.987 vs BioMèrieux VIDAS® and r = 0.972 vs. Diasorin LIAISON®, respectively. Concordance at cut-offs of 0.25 μg/l and 0.50 μg/l were 96.9% and 98.1% with Roche Cobas®, 95.4% and 96.1% with B.R.A.H.M.S Kryptor®, 93.8% and 98.4% with BioMèrieux VIDAS®, and 92.7% and 93.9% with Diasorin LIAISON®.
Conclusions:
Compared with other assays, ARCHITECT B.R.A.H.M.S PCT offers excellent precision and low-end sensitivity.
More Related Videos
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Second Order systems II
Evolutionary Relationships through Genome Comparisons
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...
Second Order systems I
By reinterpreting the system, one can derive the closed-loop transfer function, which...