Impact of image averaging on wide-field choroidal thickness measurements using enhanced-depth imaging optical
Hosein Hoseini-Yazdi1, Stephen J Vincent1, Michael J Collins1
1Contact Lens and Visual Optics Laboratory, School of Optometry and Vision Science, Queensland University of Technology, Brisbane, Queensland, Australia.
Background:
The aim of this study was to examine the influence of B-scan averaging on choroidal thickness using wide-field enhanced-depth imaging optical coherence tomography.
Methods:
Six high-resolution trans-foveal horizontal enhanced-depth imaging line scans (spanning a 60° field) were acquired consecutively from the right eye of 10 healthy adults (mean age 30 ± 5 years), with each line scan an average of 10, 20, 30, 40, 50 or 100 B-scans, using the automated real-time image averaging and follow-up features of a Spectralis device. The impact of B-scan averaging on regional measures of wide-field choroidal thickness (across macular and peripheral regions) and their accuracy was investigated, assuming that averaging 100 B-scans would provide the most accurate estimate of choroidal thickness.
Results:
Regional estimates of wide-field choroidal thickness did not vary across the different B-scan averaging conditions (all p > 0.05). The mean choroidal thickness averaged across the full wide-field area exhibited the closest agreement to measures obtained with 100 averaged B-scans, when frame averaging exceeded 30 B-scans (95 per cent limits of agreement +10 to -7, +7 to -7 and +6 to -3 μm for 30, 40 and 50 averaged B-scans, respectively), compared to 10 and 20 averaged B-scans (95 per cent limits of agreement +13 to -8 and +13 to -6 μm, respectively; p < 0.01 and p < 0.02 compared to the accuracy of 50 averaged B-scans).
Conclusion:
Averaging 30 B-scans for an individual enhanced-depth imaging optical coherence tomography line scan provided accurate measures of choroidal thickness across a wide-field (60°) area in young healthy eyes. This information can assist in designing the volumetric scan protocols required for detailed examination of the macular and peripheral choroid.
More Related Videos
09:20Novel Photoacoustic Microscopy and Optical Coherence Tomography Dual-modality Chorioretinal Imaging in Living Rabbit Eyes
Published on: February 8, 2018
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Self-Evaluation: Self-Enhancement and Self-Verification
Average Acceleration
Average Velocity