An evaluation of computed tomography dose index measurements using a pencil ionisation chamber and small detectors
Choirul Anam1, Toshioh Fujibuchi2, Freddy Haryanto3
1Department of Physics, Faculty of Mathematics and Natural Sciences, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia.
Abstract:
The aim of this study was to compare the values of the computed tomography dose index 100 (CTDI100) obtained using two small detectors (i.e. a small ionisation chamber and a small solid state detector) with those obtained from a 100 mm pencil ionisation chamber for various input CT parameters: beam width, kVp, mAs, pitch, and head-body phantom variation. The measurement of CTDI100 using the 100 mm pencil chamber was carried out in a single rotation of axial mode, while the measurement using small detectors was carried out in helical mode. The differences of CTDI100 values obtained with two small detectors were about 7% for all variations. The differences of CTDI100 values obtained with small detectors and a 100 mm pencil ionisation chamber for beam widths of more than 4 mm were within 40%. However, for the narrowest beam widths (4 mm), the difference between them was very large (about 150%).
More Related Videos
12:24Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Dose Size and Dosing Frequency: Determination Methods