Early structural valve deterioration and reoperation associated with the mitroflow aortic valve
Andrea L Axtell1,2, David C Chang3, Serguei Melnitchouk1
1Corrigan Minehan Heart Center and Division of Cardiac Surgery, Massachusetts General Hospital, Boston, Massachusetts.
Background:
Structural valve deterioration (SVD) is a known limitation of bioprosthetic valves. Recent reports have suggested a concerning rate of early SVD in patients receiving a Mitroflow aortic bioprosthesis. We therefore compared the incidence of SVD and SVD requiring reoperation among patients receiving a Mitroflow versus a common contemporary bioprosthesis.
Methods:
A retrospective cohort analysis was performed on 592 patients receiving a Mitroflow aortic bioprosthesis at our institution between 2010 and 2014. Patients were matched 1:1 using a coarsened exact matching algorithm with patients receiving a Carpentier-Edwards Magna Ease aortic bioprosthesis (Edwards Lifesciences, Irvine, CA) during the same period. The incidence of SVD (defined as a mean transprosthetic gradient ≥30 mmHg or moderate to severe intraprosthetic regurgitation), reoperation for SVD, and cumulative survival were compared between prosthesis types.
Results:
The cumulative incidence of SVD at 5 years for all patients receiving a Mitroflow aortic bioprosthesis was 16% (13-21%) and 5% underwent reoperation for SVD. Implantation of a Mitroflow valve was associated with an increased risk of SVD compared to the comparator valve (hazard ratio [HR] 2.59 [1.69-3.98], P < 0.01). Older age had a protective effect against SVD (HR 0.95 [0.93-0.96], P < 0.01). Patients who received a Mitroflow valve had reduced long-term survival compared to those who received a comparator valve (P = 0.03).
Conclusion:
The Mitroflow aortic bioprosthesis is associated with increased rates of early SVD and reoperation for valvular dysfunction as well as reduced survival compared to a contemporary valve. Enhanced clinical and echocardiographic follow-up is advisable after Mitroflow implantation.
More Related Videos
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Mortar Joint Deterioration in Masonry
The...
Aortic Regurgitation I: Introduction