Estimation of electron density, effective atomic number and stopping power ratio using dual-layer computed tomography
Shingo Ohira1, Hayate Washio2, Masashi Yagi3
1Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan; Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan.
Purpose:
Assess the accuracy for quantitative measurements of electron density relative to water (ρe/ρe,w), effective atomic number (Zeff) and stopping power ratio relative to water (SPRw) using a dual-layer computed tomography (DLCT) system.
Methods And Materials:
A tissue characterization phantom was scanned using DLCT with varying scanning parameters (i.e., tube voltage, rotation time, CTDIvol, and scanning mode) and different reference materials. Then, electron density ρe/ρe,w and atomic number Zeff images were reconstructed, and their values were determined for each reference materials. Based on these two values, SPRw was calculated. Finally, the percent error (PE) against the theoretical values was calculated for reference materials.
Results:
Significant linear relationships (p < 0.001) were observed between the measured and theoretical ρe/ρe,w (r = 1.000), Zeff (r = 0.989) and SPRw (r = 1.000) values. The PE for each reference material varied from -2.0 to 1.2% (mean, <0.1%) for electron density ρe/ρe,w, from -6.4 to 8.0% (mean, -2.0%) for atomic number Zeff, and from -2.0 to 1.9% (mean, 0.3%) for stopping power ratio SPRw. The mean PE of ρe/ρe,w (<0.1%), Zeff (<-2.5%) and SPRw (<0.4%) was verified across the variation of scanning parameters (p > 0.85).
Conclusions:
DLCT provides a reasonable accuracy in the measurements of ρe/ρe,w, Zeff and SPRw, and could enhance radiotherapy treatment planning and the subsequent outcomes.
More Related Videos
07:00Preparing Lamellae from Vitreous Biological Samples Using a Dual-Beam Scanning Electron Microscope for Cryo-Electron Tomography
Published on: August 5, 2021
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Related Concept Videos
Atomic Radii and Effective Nuclear Charge
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Configuration of Multielectron Atoms
Electronic Structure of Atoms
An atom comprises protons and neutrons, which are contained inside the dense, central core called the nucleus, with electrons present around the nucleus. Taking into account the wave–particle duality of electrons and the uncertainty in position around the nucleus, quantum mechanics provides a more accurate model for the atomic structure. It describes atomic orbitals as the regions around the nucleus where electrons of discrete energy exist, characterized by four quantum...
Atomic Orbitals
Hybridization of Atomic Orbitals I