Multi-institutional comparison of secondary check of treatment planning using computer-based independent dose
Hidenobu Tachibana1, Yukihiro Uchida2, Ryuta Miyakawa3
1Exploratory Oncology Research and Clinical Trial Center, National Cancer Center, 277-8577 Chiba, Japan; Radiation Safety and Quality Assurance Division, Hospital East, National Cancer Center, 277-8577 Chiba, Japan.
Purpose:
This report covers the first multi-institutional study of independent monitor unit (MU)/dose calculation verification for the CyberKnife, Vero4DRT, and TomoTherapy radiotherapy delivery systems.
Methods:
A total of 973 clinical treatment plans were collected from 12 institutions. Commercial software employing the Clarkson algorithm was used for verification after a measurement validation study, and the doses from the treatment planning systems (TPSs) and verification programs were compared on the basis of the mean value ± two standard deviations. The impact of heterogeneous conditions was assessed in two types of sites: non-lung and lung.
Results:
The dose difference for all locations was 0.5 ± 7.2%. There was a statistically significant difference (P < 0.01) in dose difference between non-lung (-0.3 ± 4.4%) and lung sites (3.5 ± 6.7%). Inter-institutional comparisons showed that various systematic differences were associated with the proportion of different treatment sites and heterogeneity correction.
Conclusions:
This multi-institutional comparison should help to determine the departmental action levels for CyberKnife, Vero4DRT, and TomoTherapy, as patient populations and treatment sites may vary between the modalities. An action level of ±5% could be considered for intensity-modulated radiation therapy (IMRT), non-IMRT, and volumetric modulated arc radiotherapy using these modalities in homogenous and heterogeneous conditions with a large treatment field applied to a large region of homogeneous media. There were larger systematic differences in heterogeneous conditions with a small treatment field because of differences in heterogeneity correction with the different dose calculation algorithms of the primary TPS and verification program.
Related Concept Videos
Calculating and Interpreting the Linear Correlation Coefficient
Titration Calculations: Weak Acid - Strong Base
For the titration of 25.00 mL of 0.100 M CH3CO2H with 0.100 M NaOH, the reaction can be represented as:
Titration Calculations: Strong Acid - Strong Base
A titration is carried out for 25.00 mL of 0.100 M HCl (strong acid) with 0.100 M of a strong base NaOH. The pH at different volumes of added base solution can be calculated as follows:
(a) Titrant volume = 0 mL. The solution pH is due to the acid ionization of HCl. Because this is a strong acid, the ionization is complete and the hydronium ion molarity is 0.100 M. The pH of the solution is then:
Introduction to Test of Independence
The test statistic for a test of independence is similar to that of a goodness-of-fit test:
Hypothesis Test for Test of Independence
H0: The two variables (factors)...
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...