Elham Karimi Ghodoosi1, Calogero D'Alessandria2, Ye Li3
1Institute of Radiation Protection, Helmholtz Zentrum München, German Research Center for Environmental Health, Ingolstädter Landstraße 1, 85764 Neuherberg, Germany; Nuklearmedizinische Klinik und Poliklinik, Klinikum Rechts der Isar, Technische Universität München, 81675 Munich, Germany.
Purpose:
The objective of this study was to evaluate the image degrading factors in quantitative 177Lu SPECT imaging when using both main gamma photopeak energies.
Methods:
Phantom measurements with two different vials containing various calibrated activities in air or water were performed to derive a mean calibration factor (CF) for large and small volumes of interest (VOIs). In addition, Monte Carlo simulations were utilized to investigate the effect of scatter energy window width, scatter correction method, such as effective scatter source estimation (ESSE) and triple energy window (TEW), and attenuation map on the quantification of 177Lu.
Results:
The measured mean CF using large and small VOIs in water was 4.50 ± 0.80 and 4.80 ± 0.72 cps MBq-1, respectively. Simulations showed a reference CF of 3.3 cps MBq-1 for the water-filled phantom considering all photons excluding scattered events. By using the attenuation map generated for 190 keV photons, the calculated CFs for 113 keV and 208 keV are 10% lower than by using the weighted mean energy of 175 keV for 177Lu. The calculated CF using the TEW correction was 17% higher than using the ESSE method for a water-filled phantom. However, our findings showed that an appropriate scatter window combination can reduce this difference between TEW and ESSE methods.
Conclusions:
The present work implies that choosing a suitable width of scatter energy windows can reduce uncertainties in radioactivity quantification. It is suggested to generate the attenuation map at 113 keV and 208 keV, separately. Furthermore, using small VOIs is suggested in CF calculation.
Related Concept Videos
Calculating Standard Free Energy Changes
Calculation of Volume of Solids by Integration
Conservation of Energy in Control Volume
For steady flow systems, the time derivative of the stored energy becomes zero since there is no energy accumulation within the control volume. This simplifies the energy equation to:
Transcription Attenuation in Prokaryotes
There are several different mechanisms used to attenuate transcription. In ribosome mediated...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...