Arm and finger measurements in the third trimester: Implications for blood pressure measurement
Victoria A Eley1, Leonie Roberts1, Leah Rickards2
1The Royal Brisbane and Women's Hospital, Butterfield St, Herston, 4006 Queensland, Australia; The University of Queensland, St Lucia 4067, Queensland, Australia.
Objectives:
To obtain arm and finger measurements of women ≥32 weeks gestation to determine: the requirement for different arm cuff sizes; the suitability of available finger cuffs in this population; the best predictor of arm conicity; the frequency of cuff placement on the forearm or leg.
Study Design:
Prospective observational pilot study.
Main Outcome Measures:
Right and left mid-arm circumference (MAC) and to compare these to the recommended cuff sizes; right and left finger circumference; right and left arm conicity; the responses of women to a three-point Likert scale regarding cuff placement.
Results:
Measurements were obtained for 450 women at an Australian tertiary hospital with a median (IQR) gestation of 35.7 (34.0-37.0); 299 (66.4%) were Caucasian and 35 (7.8%) had gestational hypertension. The median (IQR) body mass index (BMI) was 29.6 kg/m2 (26.2-33.4), range 18.0-62.2. Median (IQR) right MAC was 29.9 cm (27.4-33), range 19.6-53.2. Based on right MAC, 58 (12.9%) required a large cuff and 6 (1.3%) a thigh cuff. Maximum right finger circumference was 7.0 cm. BMI, weight and right MAC were positively correlated with right arm conicity: r = 0.51, 0.42 and 0.45, p < 0.001 for all. R2 for each were 0.26, 0.17 and 0.20. Fourteen (3.1%) reported cuff placement on the forearm or leg.
Conclusions:
A small percentage of women are likely to be unsuited to traditional arm cuffs. Available finger-cuffs would suit this population. BMI could potentially be used to select women with cone-shaped arms for future studies of alternative devices.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: