Thoracic endografting increases cardiac afterload and leads to left ventricular hypertrophy in dogs
Yoshiyuki Yamashita1, Yasuhisa Oishi1, Yuma Motomatsu1
1Department of Cardiovascular Surgery, Kyushu University Graduate School of Medical Sciences, Fukuoka, Japan.
Objectives:
Aortic endografting can cause aortic stiffening. We aimed to determine the chronic effect of thoracic endografting on cardiac afterload, function and remodelling.
Methods:
Eleven dogs were included, and all except 1 was successfully assessed [endograft, n = 5; sham operation (control), n = 5]. We deployed a stent graft in the descending aorta. The ascending aortic pressure and flow were measured, and aortic input impedance was obtained by frequency analysis to determine characteristic impedance and arterial compliance. Left ventricular pressure-volume relations were measured with an admittance catheter. Measurements were performed before, 10 min after and 3 months after endografting. Following euthanasia, we weighed the left ventricle of each dog and measured the cardiomyocyte cell size.
Results:
Arterial compliance decreased from 0.47 ± 0.07 to 0.36 ± 0.06 and to 0.31 ± 0.05 ml/mmHg (both P < 0.01 versus baseline), and characteristic impedance increased from 0.11 ± 0.04 to 0.19 ± 0.05 and to 0.21 ± 0.04 mmHg/ml/s (both P < 0.01 versus baseline) 10 min and 3 months after endografting, respectively. Pressure-volume relation analysis showed that arterial elastance increased from 5.3 ± 1.0 to 6.7 ± 1.6 (at 10 min) and to 6.8 ± 1.0 mmHg/ml (at 3 months) (both P < 0.05 versus baseline), but end-systolic elastance and ventriculo-arterial coupling remained unchanged. Left ventricular weight to body weight ratio and left ventricular cardiomyocyte cell width in the endograft group were larger compared with the control's results (5.06 ± 0.27 g/kg vs 4.20 ± 0.49 g/kg, P = 0.009, 15.1 ± 1.7 µm vs 13.9 ± 1.5 µm, P = 0.02, respectively).
Conclusions:
The mid-term effect of the descending aortic endografting on left ventricular contractility and efficiency in canine normal hearts was minimal. However, endografting resulted in increased cardiac afterload and left ventricular hypertrophy.
More Related Videos
08:34Technique of Minimally Invasive Transverse Aortic Constriction in Mice for Induction of Left Ventricular Hypertrophy
Published on: September 25, 2017
06:47A Pulmonary Trunk Banding Model of Pressure Overload Induced Right Ventricular Hypertrophy and Failure
Published on: November 29, 2018
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Thoracic Aorta
Increasing Function
The Thoracic Cage: Sternum
The sternum is the elongated bony structure on the anterior side of the thoracic cage. It consists of three parts: the manubrium, the body, and the xiphoid...
The Thoracic Cage: Ribs
Parts of a Typical Rib
A typical rib has a head, neck, and body. The posterior end of the rib is called the head, followed by a narrow neck. The head articulates primarily with the costal...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...