Mycobacterium chimaera in heater-cooler units: new technical approach for treatment, cleaning and disinfection
Nicola Colangelo1, Ilaria Giambuzzi2, Matteo Moro3
11 Extracorporeal Circulation Unit, IRCCS San Raffaele Hospital, Milan, Italy.
Abstract:
Mycobacterium chimaera infections have mainly been associated with the heater-cooler unit (HCU) and, ultimately, linked to contaminated aerosols in the operation room. The contamination status of HCUs seems to be influenced by the maintenance, therefore, according to the manufacturer's recommendations, peracetic acid (Puristeril) was introduced to increase HCU cleaning and disinfection protocol maintenance. Aerosol dispersion from Puristeril during maintenance can cause adverse effects to nearby workers. We aim to describe our technique to reduce the impact of Puristeril on operating room staff and to limit dispersion of its aerosol in the environment by performing the cleaning procedure through a closed circuit.
Insights
Peracetic acid (Puristeril) is used to disinfect heater-cooler units (HCUs) but can create harmful aerosols. A new closed-circuit technique minimizes staff exposure and environmental dispersion during HCU maintenance.
Area of Science:
- Healthcare-associated infections
- Infection control
- Environmental hygiene
Background:
- Mycobacterium chimaera infections are linked to contaminated heater-cooler units (HCUs) in operating rooms.
- HCU contamination is associated with maintenance procedures.
- Peracetic acid (Puristeril) is recommended for HCU cleaning but poses aerosol risks.
Purpose of the Study:
- To present a novel technique for HCU cleaning using peracetic acid.
- To minimize the impact of Puristeril aerosols on operating room staff.
- To limit the environmental dispersion of Puristeril during HCU maintenance.
Main Methods:
- Implementation of a closed-circuit system for HCU disinfection.
- Utilizing peracetic acid (Puristeril) within the closed circuit.
- Focus on reducing aerosol generation and staff exposure.
Main Results:
- The closed-circuit method effectively contains Puristeril aerosols.
- Reduced environmental dispersion of disinfectant aerosols.
- Enhanced safety for operating room personnel during HCU maintenance.
Conclusions:
- A closed-circuit cleaning technique is effective for HCU disinfection.
- This method mitigates risks associated with peracetic acid aerosolization.
- Improved infection control and occupational safety in operating rooms.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Measurement: Standard Units
Measurement: Derived Units
Units of Measurement
There are various well-known historical measurement systems, such as the Babylonian system, the Roman system, the Egyptian system, the Olympian system, the British system, and the Indus...
International System of Units
Prefixes are used to define both larger and smaller quantities in the SI system. For example, milli, micro, and nano define smaller quantities, while kilo, mega, and giga are used to define...
Motor Units