Comprehensive Evaluation of Compendial USP, BacT/Alert Dual-T, and Bactec FX for Detection of Product Sterility
Matthew R England1, Frida Stock1, James E T Gebo1
1Department of Laboratory Medicine, Clinical Center, National Institutes of Health, Bethesda, Maryland, USA.
Abstract:
The emergence of cell therapy programs in large academic centers has led to an increasing demand for clinical laboratories to assist with product sterility testing. Automated blood culture systems have shown promise as alternatives to the manual USP<71> compendial method, but current published data are limited by small organism test sets, particularly for molds. In 2015, failure of the Bactec FX system to detect mold contamination in two products prompted us to evaluate three test systems (compendial USP<71>, Bactec FX, and BacT/Alert Dual-T) over seven different culture combinations, using 118 challenge organisms representative of the NIH current good manufacturing practice (cGMP) environment. At <96 h and <144 h for bacterial and fungal detection, respectively, the compendial USP<71> method significantly outperformed the Bactec FX system (84.7% versus 64.4%; P = 0.0006) but not the BacT/Alert system at 32.5°C (78.8%; P = 0.3116). Extended incubation to 360 h with terminal visual inspection improved sensitivity, without a significant difference between compendial USP<71> and BacT/Alert testing (95.7% versus 89.0%; P = 0.0860); both systems were better than the Bactec FX system (71.2%; P < 0.0001 and P = 0.0003, respectively). The Bactec FX and BacT/Alert systems performed equivalently for 30 isolates derived from clinical bloodstream infections, confirming system optimization for clinical organisms rather than environmental contaminants. Paired Sabouraud dextrose agar (SDA) plates were always positive for fungi within the acceptable time frame. This study shows that the Bactec FX system is suboptimal for product sterility testing, and it provides strong data to support the use of BacT/Alert testing at 32.5°C paired with a supplemental SDA plate as an acceptable alternative to the compendial USP<71> method for product sterility testing.
More Related Videos
Related Concept Videos
In Vitro Drug Dissolution: Compendial Testing Models I
In Vitro Drug Dissolution: Compendial Testing Models II
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...