Starfix lead extraction: Clinical experience and technical issues
Pier Giorgio Golzio1, Ilaria Meynet1, Fulvio Orzan1
1Division of Cardiology, Department of Internal Medicine, Azienda Ospedaliero-Universitaria Città della Salute e della Scienza di Torino, Turin, Italy.
Abstract:
Transvenous lead extraction (TLE) of the Starfix coronary sinus (CS) active-fixation lead may be challenging, due to undeployment of fixation lobes and venous occlusion. We report our experience in Starfix TLE, in comparison with previous data. A 78-year-old male, implanted in 2009 with Starfix lead, was referred to our institution for TLE, due to infective endocarditis with lead-associated vegetations. The tip of Starfix lead was located in distant, anterior position, in the great cardiac vein, close to patent left internal mammary artery-to-left anterior descending artery anastomosis, and first-choice surgical removal had a prohibitive operative risk. Conventional dilatation beyond CS ostium, as well as the use of a standard delivery catheter, was ineffective. An off-label modification of the delivery, by cutting the distal soft tip, was successful. However, the tip of the lead fragmented and was trapped in the innominate vein. Then a gooseneck snare grasped the fragment, allowing complete retrieval. TLE of Starfix leads may be particularly challenging, especially when its tip is located in a distant anterior location. In these cases, important help may be obtained by dilatation within the CS, by means of conventional or modified delivery catheters. Only experienced operators, sometimes with non-conventional techniques, should perform TLE of Starfix leads. <Learning objective: TLE of Starfix leads may be challenging, particularly when the tip is located in a distant anterior position. Dilatation with conventional tools may be precluded. In these cases modifications of the delivery catheters may be useful. Surgery should be avoided as first-choice procedure; only experienced operators, sometimes with non-conventional techniques, should perform TLE of Starfix leads.>.
Related Concept Videos
Ethical Issues
Ethical Concerns in Healthcare:
Issues And Trends In Healthcare Delivery System
Cost Containment
Payment for healthcare services has historically promoted adoption of costly and often unnecessary or inefficient...
The Stanford Prison Experiment
What is an Experiment?
Phase-lead and Phase-lag Controllers
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...