Relation of left atrial appendage closure devices to topographic neighboring structures using standardized imaging by
Simon Lindner1, Michael Behnes1, Annika Wenke1
1First Department of Medicine, University Medical Center Mannheim (UMM), Faculty of Medicine Mannheim, University of Heidelberg, European Center for AngioScience (ECAS), and DZHK (German Center for Cardiovascular Research), Mannheim, Germany.
Background:
Although left atrial appendage (LAA) anatomy and topographic relations are well understood, little is known about the impairment of neighboring structures (NBS) by an implanted left atrial appendage closure (LAAC) device. This prospective longitudinal observational study for the first time describes distances of implanted LAA closure (LAAC) devices to NBS using a standardized imaging protocol of cardiac computed tomography angiography (cCTA).
Hypothesis:
cCTA imaging is an eligible tool for post-implantation evaluation of LAAC devices and their relation to neighboring structures.
Methods:
cCTA data sets of consecutive patients 6 months after successful LAAC were acquired on a third generation dual-source CT system and reconstructed with a slice thickness of 0.5 mm. The standardized multi-planar reconstruction LAA occluder view for post-implantation evaluation (LOVE) algorithm was used to measure the distances to NBS in relation to LAA morphology and implanted LAAC devices.
Results:
A total of 48 patients (median age 80 years, 25% female) were included. Left upper pulmonary vein and circumflex artery were generally closest to occlusion devices (median 2.9 and 2.8 mm, respectively). AMPLATZER AMULET devices were closer to the mitral valve annulus than WATCHMAN devices (6.6 mm (inter quartile range [IQR] 4.9-8.6) vs 10.9 mm (IQR 7.4-14.0), P = 0.001). Distances to the left upper pulmonary vein were affected by LAA morphology, with cauliflower type having the closest proximity (1.7 mm [IQR 1.0-3.4], P = 0.048).
Conclusion:
A standardized cCTA imaging protocol is an eligible tool to accurately measure distances to NBS. Left upper pulmonary vein and circumflex artery are closest to LAAC devices and could thus be most prone to impairment.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Structure of Cardiac Muscles
Compared to skeletal muscles, cardiac muscle cells are small and mostly have a single nucleus. Additionally, they are usually...
Cardiac Catheterization III: Left Heart Catheterization
Surface Appendages of Archaea
Topographic Surveying and Contours