Transient power elevation during iron dextran infusion in a patient with a continuous-flow left ventricular assist
Paulino A Alvarez1, Nicole Byram2, J Brad Williams3
11 Cardiology Division, Department of Internal Medicine, University of Iowa, Iowa City, IA, USA.
Abstract:
Anemia is common in patients with mechanical circulatory support and is associated with increased morbidity. Repletion using parenteral iron infusions has been proven to be beneficial in patients with heart failure. In this report, we describe a case of increased power and flows of continuous-flow left ventricular assist device (LVAD) during an iron dextran infusion. We subsequently studied the effects of iron dextran infusion in an in vitro LVAD mock circulatory loop. The observed increase in flow and power was most likely due to drug-patient interaction rather than drug-LVAD interaction. Mock loops and in vivo animal models may be necessary for proactive evaluation of the safety of intravenous (IV) preparations in this patient population.
Related Concept Videos
The Power Flow Problem and Solution
Control of Power Flow
Maximum Power Flow and Line Loadability
Freezing Point Depression and Boiling Point Elevation
The boiling point of a liquid is the temperature at which its vapor pressure is equal to ambient atmospheric pressure. Since the vapor pressure of a solution is lowered due to the presence of nonvolatile solutes, it stands to reason that the solution’s boiling point will subsequently be increased. Vapor pressure increases with temperature, and so a solution will require a higher temperature than will pure solvent to achieve any given vapor pressure, including one...
Power
Elevation of Intermediate Points on Vertical Curves