Transaxillary transcatheter aortic valve implantation utilizing a novel vascular closure device with resorbable
Andreas Schaefer1, Johannes Schirmer2, Niklas Schofer3
1Department of Cardiovascular Surgery, University Heart Center Hamburg, Martinistrasse 52, 20246, Hamburg, Germany. and.schaefer@uke.de.
Objectives:
We herein aimed to evaluate technical feasibility of transaxillary (Tax) transcatheter aortic valve implantation (TAVI) utilizing a novel vascular closure device with a resorbable collagen plug and absence of suture material.
Methods:
Between 05/2018 and 8/2018, eight patients (76.0 ± 5.9 years, 62.5% male, logEuroSCORE I 23.6 ± 4.7) received Tax-TAVI using the MANTA™ vascular closure device. Implanted transcatheter heart valves consisted of Edwards Sapien 3, NVT Allegra, Medtronic CoreValve EvolutR and SJM Portico.
Results:
Puncture location depth was variable (3.5-7.5 cm). The left subclavian artery was used in five cases, the right subclavian artery in three cases. Low-pressure balloon-angioplasty for vessel closure was performed in 5/8 patients. VARC-2 defined device success was met in all patients. Major access site complication occurred in one patient with aneurysma spurium of the subclavian artery and consecutive stent implantation on postoperative day 5.
Conclusion:
The MANTA™ device is applicable in Tax-TAVI, with potential particular advantages regarding easiness of use and marked access for subsequent interventions in case of vascular complications. Before conclusions regarding clinical efficacy and safety can be made, the device has to be evaluated in larger patient cohorts.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Related Concept Videos
Seedless Vascular Plants
Imaging Studies VII: Vascular Imaging
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
Angle Closure Glaucoma: Treatment
Members Made of Elastoplastic Material
As the bending moment...