Related Experiment Video
Updated: Jan 31, 2026

Real-time Monitoring of Mitochondrial Respiration in Cytokine-differentiated Human Primary T Cells
Published on: October 19, 2021
Monitoring respiration using the pressure sensors in a dialysis machine
Frida Sandberg1, Mattias Holmer2, Bo Olde2
1Department of Biomedical Engineering, Lund University, Lund, Sweden.
Objective:
Although respiratory problems are common among patients with end-stage renal disease, respiration is not continuously monitored during dialysis. The purpose of the present study is to investigate the feasibility of monitoring respiration using the pressure sensors of the dialysis machine.
Approach:
Respiration induces variations in the blood pressure that propagates to the extracorporeal circuit of the dialysis machine. However, the magnitude of these variations are very small compared to pressure variations induced by the dialysis machine. We propose a new method, which involves adaptive template subtraction and peak conditioned spectral averaging, to estimate respiration rate from the pressure sensor signals. Using this method, an estimate of the respiration rate is obtained every 5th second provided that the signal quality is sufficient. The method is evaluated for continuous monitoring of respiration rate in nine dialysis treatment sessions.
Main Results:
The median absolute deviation between the estimated respiration rate from the pressure sensor signals and a reference capnography recording was 0.02 Hz (1.3 breaths per min).
Significance:
Our results suggest that continuous monitoring of respiration using the pressure sensors of the dialysis machine is feasible. The main advantage with such monitoring is that no additional sensors are required which may cause patient discomfort.
More Related Videos
11:06A Human-machine-interface Integrating Low-cost Sensors with a Neuromuscular Electrical Stimulation System for Post-stroke Balance Rehabilitation
Published on: April 12, 2016
08:31The Calibration and Use of Capacitance Sensors to Monitor Stem Water Content in Trees
Published on: December 27, 2017
Related Concept Videos
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Dialysis
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Respiration
Energy production in the human body is primarily fueled by oxidation, a process where food molecules are burned by combining with oxygen to produce carbon dioxide and water. This vital metabolic process sustains life, and is supported intricately by the respiratory system.
Structure and Function of the Respiratory System:
The respiratory system is a complex network of structures that includes the nose, oropharynx, larynx, trachea,...
Errors occurring during blood pressure monitoring
Several factors...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...