Symmetry Matching of the Medial Acetabular Surface-A Quantitative Analysis in View of Patient-Specific Implants
Georg Osterhoff1,2, Andreas Petersik3, Kai Sprengel1
1Department of Trauma, University Hospital Zurich, Zürich, Switzerland.
Objective:
To quantify intrapelvic surface symmetry in reference to a preshaped suprapectineal acetabular implant.
Methods:
In this cross-sectional study, an anatomically preshaped acetabular fracture implant was fitted on 3D surface models of 516 pelvises from a preexisting bone database using a software tool for automated implant fitting (SOMA, Stryker Orthopaedic Modeling and Analytics) of a CAD model of the implant. The distances between bone and the reference implant were measured at 2310 reference points for each hemipelvis.
Results:
The average distance between the left hemipelvis and the plate was 1.98 mm (median, 10% percentile: 1.45, 90% percentile: 2.78) and 2.0 mm (median, 10% percentile: 1.45, 90% percentile: 2.92) between the right hemipelvis and the plate. There was no significant difference between the 2 hemipelvises (median absolute pairwise delta: 0.25 mm; 10% percentile: 0.04, 90% percentile: 0.82; Wilcoxon, P = 0.064).
Conclusions:
With regard to the periacetabular surface of the inner pelvis, the pelvis can be considered sufficiently symmetric for using the mirrored contralateral hemipelvis as a template for patient-specific implants in acetabular fracture fixation.
Related Concept Videos
Symmetry
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Sign Test for Matched Pairs
To conduct the sign test, we first calculate the differences in...
Gauss's Law: Planar Symmetry
Symmetry in Maxwell's Equations
Econometric Views (EViews)