Published on: September 11, 2011
Reduction of Metal Artifacts and Improvement in Dose Efficiency Using Photon-Counting Detector Computed Tomography
Wei Zhou, David J Bartlett, Felix E Diehn
1Health Sciences Research, Mayo Clinic, Jacksonville, FL.
Objectives:
The aim of this study was to investigate the impact on metal artifacts and dose efficiency of using a tin filter in combination with high-energy threshold (TH) images of a photon-counting detector (PCD) computed tomography (CT) system.
Materials And Methods:
A 3D-printed spine with pedicle screws was scanned on a PCD-CT system with and without tin filtration. Image noise and severity of artifacts were measured for low-energy threshold (TL) and TH images. In a prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant study, 20 patients having a clinical energy-integrating detector (EID) CT were scanned on a PCD-CT system using tin filtration. Images were reviewed by 3 radiologists to evaluate visualization of anatomic structures, diagnostic confidence, and image preference. Artifact severity and image noise were measured. Wilcoxon signed rank was used to test differences between PCD-CT TH and EID-CT images.
Results:
Phantom TH images with tin filtration reduced metal artifacts and had comparable noise (32 HU) to TL images (29 HU) acquired without tin filtration. Visualization scores for the cortex, trabeculae, and implant-trabecular interface from PCD-CT TH images (4.4 ± 0.9, 4.4 ± 1.0, and 4.4 ± 1.0) were significantly higher (P < 0.0001) than EID-CT images (3.3 ± 1.3, 3.3 ± 1.2, and 3.3 ± 1.6). A strong preference was shown for PCD-CT TH images due to improved diagnostic confidence and decreased artifact severity. Noise in PCD-CT TH images (93 ± 41 HU) was significantly lower than that in EID-CT images (133 ± 92 HU, P < 0.05).
Conclusions:
Threshold high images acquired with tin filtration on PCD-CT demonstrated a substantial decrease in metal artifacts and an increase in dose efficiency compared with EID-CT.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Oxidation-Reduction Reactions
Filtration and Urine Formation
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Glomerular Filtration: Net Filtration Pressure
GBHP, with an average value of 55 mmHg, promotes filtration by pushing water and solutes through the filtration membrane. This is balanced by two opposing forces: CHP, a "back pressure" exerted against the filtration membrane by fluid already in the capsular space and renal...