A novel magnetic resonance elastography transducer concept based on a rotational eccentric mass: preliminary
Jurgen Henk Runge1, Stefan Heinz Hoelzl, Jelizaveta Sudakova
1Division of Imaging Sciences and Biomedical Engineering, King's College London, London, United Kingdom. Department of Radiology and Nuclear Medicine, Amsterdam UMC, location AMC, Amsterdam, The Netherlands. Author to whom any correspondence should be addressed.
Background:
Magnetic resonance elastography (MRE) is used to non-invasively estimate biomechanical tissue properties via the imaging of propagating mechanical shear waves. Several factors including mechanical transducer design, MRI sequence design and viscoelastic reconstruction influence data quality and hence the reliability of the derived biomechanical properties.
Purpose:
To design and characterize a novel mechanical MRE transducer concept based on a rotational eccentric mass, coined the gravitational transducer.
Materials And Methods:
Table top measurements were performed using accelerometers to characterize the frequency response of the new transducer concept at different driving frequencies (f VIB) and different rotating masses. These were compared to a commercially available pneumatically driven MRE transducer. MR data were acquired on a 3T scanner using a fractionally encoded gradient echo MRE sequence in three healthy volunteers. Acceleration and displacement spectra were plotted in units of g and mm, respectively, and visually compared, emphasizing the ratio between the peaks at f VIB and its 2nd harmonic, a known cause of error in the reconstruction of biomechanical properties as is explored in more detail in numerical simulations here. No formal statistical testing was performed in this proof-of-principle paper.
Results:
The new transducer concept shows-as expected from theory-a quadratic or linear increase of acceleration amplitude with increase in f VIB or mass, respectively. Furthermore, different versions of the transducer show markedly lower 2nd harmonic-to-f VIB ratios compared to the commercially available pneumatically driven transducer. Displacement was constant over a range of f VIB, in accordance with theory. Phantom and in vivo data show low nonlinearity and excellent data quality.
Conclusion:
The table top measurements are in concordance with the theory behind a transducer based on a rotational eccentric mass. The resulting constant displacement amplitude irrespective of f VIB and low 2nd harmonic-to-f VIB ratio result in low nonlinearity and high data fidelity in both phantom and in vivo examples.
More Related Videos
12:18Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
07:02Investigating the Potential of Singly Curved Thin Piezoelectric Transducers for Energy Harvesting and Structural Health Monitoring
Published on: November 14, 2025
Related Concept Videos
Formula Mass and Mole Concepts of Compounds
Transducer Mechanism: Nuclear Receptors
About 48 different soluble family members of nuclear receptors are identified that can be divided into two main classes:
Transducer Mechanism: G Protein–Coupled Receptors
GPCRs are also called heptahelical,...
Transducer Mechanism: Enzyme-Linked Receptors
Major types that are helpful drug targets include:
Concept of Resonance and its Characteristics
Gravitation Between Spherically Symmetric Masses