[Field Shape Optimization Technique Based on Dose Volume Histogram Using Daily Cone-beam Computed Tomography in
Yuto Kimura1, Suguru Dobashi2, Yoshiki Ishizawa2
1Course of Radiological Technology, Health Sciences, Tohoku University School of Medicine (Current address: Division of Medical Physics, Aomori Shintoshi Hospital).
Abstract:
This study aimed to develop and evaluate field shape optimization technique based on dose calculation using daily cone-beam computed tomography (CBCT) to compensate for interfractional anatomic changes in three-dimensional conformal radiation therapy (3D-CRT) for prostate cancer. For each of 10 patients, 9-10 CBCT images were obtained throughout the treatment course. The prostate, seminal vesicles, and rectum were manually contoured in all CBCT images. Subsequently, plan adaptation was performed with a program developed in-house. This program calculates dose distributions on CBCT images and optimizes field shape to minimize rectal dose while keeping the target at the optimal dose coverage (the planning target volume D95% receives 95% of the prescription dose). To evaluate the adaptive planning approach, we re-calculated dose distributions on CBCT images based on the conventional and adaptive plans. For the entire cohort, plan adaptation improved rectal V50 Gy, V60 Gy, V65 Gy, and V70 Gy by -7.71±8.43%, -8.30±8.90%, -7.91±8.51% and -7.03±7.70% on average (±SD), respectively. Our results demonstrate that adaptive planning approach is superior to the conventional planning approach for optimizing dose distribution, and this adaptive approach can optimize field shape in 3 min. The proposed approach can be an effective solution for the problem of interfractional anatomic changes in 3D-CRT for prostate cancer.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Conformity
Biological Effects of Radiation
Cancer Therapies
However, cancer treatments can pose several challenges, as therapies used to kill cancer cells are generally also toxic to normal cells. Moreover, cancer cells mutate rapidly and can develop resistance to chemical agents or radiation therapy. Besides, all types of cancer cells may not respond to the same therapy. Some cancer cells respond to one...
Imaging Studies III: Computed Tomography
Probability Histograms