Troponin T release comparing manually or magnetically guided radiofrequency ablation for AVNRT- a MAGMA AVNRT
Ester Anne Kringeland1, Isabel Deisenhofer2, Gabriele Hessling2
1a Department of Cardiology , Haukeland University Hospital , Bergen , Norway.
Objectives:
Catheter ablation is regarded as first-line therapy for symptomatic atrioventricular nodal reentry tachycardia (AVNRT). Ablation induces intended myocardial damage and the extent of myocardial damage may differ between ablation methods. The objective of this MAGMA AVNRT(NCT00875914) substudy was to compare high-sensitive cardiac troponin T (hs-cTnT) levels as a surrogate marker for myocardial damage after manually guided (MAN) AVNRT ablation versus AVNRT ablation using remote magnetic navigation (RMN).
Design:
In total, 70 patients (mean age 44 ± 14 years, 26% male) undergoing catheter ablation for AVNRT in the MagMa-AVNRT-Trial were randomized to remote magnetic navigation (n = 34, 49%) or manually guided catheter ablation (n = 36, 51%). hs-cTnT was measured the day after the procedure.
Results:
The median follow-up time was 6.2 ± 1.1 years. Acute success was 100% in both groups. hs-cTnT release was significantly lower in the remote magnetic navigation group (52 ng/L versus 95 ng/L, p < .01), even though the ablation time was longer and number of applications was higher with remote magnetic navigation (4.2 min vs 2.8 min, p = .017; 4.9 vs 3.3 applications, p = .01). hs-cTnT released per minute ablation time was also lower with remote magnetic navigation (12 ng/L versus 34 ng/L, p < .01). Both groups exhibited similar clinical long-term follow up regarding recurrence and complications.
Conclusion:
Remote magnetic navigation controlled catheter ablation of AVNRT has similar clinical outcome, but leads to less hs-cTnT release than manually guided catheter ablation. This might correspond to less unintended myocardial damage with RMN, which might be advantageous in complex ablation procedures.
More Related Videos
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Related Concept Videos
Comparative Excretory Systems
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Energy-releasing Steps of Glycolysis
The first energy-releasing step—the 6th step of glycolysis...
Comparing Experimental Results: Student's t-Test