Technical Note: Use of commercial multilayer Faraday cup for offline daily beam range verification at the McLaren
Biniam Tesfamicael1, Basit Athar1, Ana Bejarano Buele1
1Karmanos Cancer Institute, McLaren Proton Therapy Center, Flint, MI, USA.
Purpose:
Daily verification of the proton beam range in proton radiation therapy is a vital part of the quality assurance (QA) program. The objective of this work is to study the use of a multilayer Faraday cup (MLFC) to perform a quick and precise daily range verification of proton beams produced by a synchrotron.
Methods:
Proton beam depth dose measurements were performed at room iso-center in water using PTW water tank and Bragg Peak ion chamber. The IBA Giraffe, calibrated against the water tank data, was used to measure the water equivalent thickness (WET) of the sample copper plates. The WET measurements provided the range calibration factors for the MLFC. To establish a baseline for in room measurements, range measurements for energies from 70 to 250 MeV in steps of 10 MeV were performed using the Pyramid MLFC at room iso-center. For the daily range verification measurements, the MLFC is permanently placed at the end of the beam line, inside the accelerator vault. The daily range constancy is performed for five representative beam energies; namely 70, 100, 150, 200, and 250 MeV. Data collected over a period of more than 100 days are analyzed and presented.
Results:
The measured WET values of the copper plates increased with increasing energy. The centroid channel number in the MLFC where the protons stop, was converted to depth in water and compared to the depth of the distal 80% (d80) obtained from the water tank measurements. The depths agreed to within 2 mm, with the maximum deviation of 1.97 mm observed for 250 MeV beam. The daily variation in the ranges measured by the MLFC was within ±0.5 mm. The total time to verify five proton beam ranges varies between 4 and 5 min.
Conclusion:
Based on the result of our measurements, the MLFC can be used for a daily range constancy check with submillimeter accuracy. It is a quick and simple method to perform range constancy verification on a daily basis.
More Related Videos
04:03Author Spotlight: Evaluating the Therapeutic Efficacy of Moving Cupping Along Meridians for Acute Exacerbation of COPD
Published on: September 27, 2024
08:16Method for the Assessment of Effects of a Range of Wavelengths and Intensities of Red/near-infrared Light Therapy on Oxidative Stress In Vitro
Published on: March 21, 2015
Related Concept Videos
Faraday's Law
Self-Evaluation: Self-Enhancement and Self-Verification
Faraday Disk Dynamo
Strategies of Self-Presentation II: Self-Verification
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...