Standardization of a Cytometric Bead Assay Based on Egg-Yolk Antibodies
Published on: May 19, 2023
Standardization of the HbA2 Assay
Renata Paleari1, Andrea Mosca1
1Department of Physiopathology and Transplantation, Center for Metrological Traceability in Laboratory Medicine (CIRME), University of Milano, Italy.
Background:
A project for the standardization of HbA2 was launched by the IFCC back in 2004.
Materials And Methods:
In this work we report on the state-of-the-art of the project on standardization of HbA2. Data obtained from various EQAS studies, and from previous experimental evaluations, are presented.
Results:
We have proven that biases between various commercial methods are still currently significant. We have also shown that calibration by commutable control materials may halve the inter-method variability.
Conclusions:
The foundation of the reference system for HbA2, together with a brief preliminary presentation of the proposed primary reference measurement procedure based on ID-MS are outlined.
More Related Videos
07:28Standardized In vitro Assays to Visualize and Quantify Interactions between Human Neutrophils and Staphylococcus aureus Biofilms
Published on: June 8, 2022
13:58Amplifying and Quantifying HIV-1 RNA in HIV Infected Individuals with Viral Loads Below the Limit of Detection by Standard Clinical Assays
Published on: September 26, 2011
Related Concept Videos
Standard Enthalpy of Formation
Standard Electrode Potentials
Measurement: Standard Units
Calculating Standard Free Energy Changes
Standard Solutions
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...