Rate and predictors of electrical failure in non-recalled defibrillator leads
Furqan Khattak1, Aman Gupta1, Krishna Alluri1
1Heart and Vascular Institute, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
Background:
Implantable cardioverter defibrillator (ICD) leads are considered as the 'weakest link' in defibrillator systems due to FDA recalls and advisories involving popular lead models from major manufacturers. The rate of electrical failure of ICD leads not implicated in a recall is however not well determined.
Methods:
Medical records of patients implanted with ICDs at hospitals of the University of Pittsburgh Medical Center between 2002 and 2014 were analyzed. Leads were classified as having electrically failed if removed or replaced for reasons other than infection or heart transplantation. Patients were followed to endpoint of death or electrical lead failure.
Results:
2410 consecutive ICD recipients (mean age 66 ± 13 years, women 22%, single/dual/biventricular-ICD 20%/44%/36%) were included. During a mean follow-up of 3.9 ± 3.3 years, 1272 patients (53%) died, 55 patients (2.3%) had ICD lead electrical failure, and 1052 (44%) patients were alive with functional leads at the time of last follow-up. Patients with failed leads had higher BMI (p = 0.07), better functional status (p = 0.04), higher serum creatinine (p = 0.004), wider QRS complex (p = 0.01), higher number of implanted leads (p = 0.06) and were more likely to have ischemic cardiomyopathy (p = 0.03). After adjusting for these variables in a binary logistic regression model, only a lower BMI, presence of non-ischemic cardiomyopathy, and a better functional status remained independently predictive of electrical failure.
Conclusions:
Only 2.3% of non-recalled ICD leads experience electrical failure (annual failure rate of 0.6%). A higher patient functional status, lower BMI, and non-ischemic etiology of cardiomyopathy are independently associated with higher rates of ICD lead failure.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:35Examining Recall Memory in Infancy and Early Childhood Using the Elicited Imitation Paradigm
Published on: April 28, 2016
Related Concept Videos
Measuring Reaction Rates
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...
Temperature Dependence on Reaction Rate
Atoms, molecules, or ions must collide before they can react with each other. Atoms must be close together to form chemical bonds. This premise is the basis for a theory that explains many observations regarding chemical kinetics, including factors affecting reaction rates.
The collision theory is based on the postulates that (i) the reaction rate is proportional to the rate of reactant collisions, (ii) the reacting species collide in an orientation allowing contact between...
Phase-lead and Phase-lag Controllers
Related Rates
Speciation Rates