Generic gas chromatography flame ionization detection method using hydrogen as the carrier gas for the analysis of
Frank Bernardoni1, Holst M Halsey1, Robert Hartman1
1Process Research & Development, MRL, Merck & Co., Inc., Rahway, NJ, 07065, USA.
Abstract:
Within the pharmaceutical industry, the determination of residual solvents by Gas Chromatography Flame Ionization Detection (GC-FID) is a highly utilized analytical test that often employs helium (He) as the carrier gas. However, many do not realize that helium is a non-renewable resource that will eventually become progressively more difficult to source. In recent years, analytical chemists are increasingly adopting hydrogen (H2) in place of helium for routine GC analysis. In this study, a simple and efficient generic/universal GC-FID method using H2 as the carrier gas has been developed with the capability of baseline resolution of over 30 of the most commonly used solvents in development and manufacturing with a method run time of less than eight minutes. The use of this method for the separation and analysis of solvents within a pharmaceutical manufacturing process is demonstrated with additional method validation data presented using five different diluents as a means to increase flexibility for the chromatographer. Furthermore, it is the recommendation of the authors that the current compendia for residual solvent analysis be updated to allow for hydrogen as a carrier gas. The similarity between He and H2 observed within this study supports the use of hydrogen as a suitable replacement for helium, and an update of the EU and USP compendia for residual solvent analysis should be made to reflect this.
More Related Videos
08:22Nitrogen Compound Characterization in Fuels by Multidimensional Gas Chromatography
Published on: May 15, 2020
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....