Correction of QCT vBMD using MRI measurements of marrow adipose tissue
Xiaoguang Cheng1, Glen M Blake2, Zhe Guo1
1Department of Radiology, Beijing Jishuitan Hospital, Beijing 100035, China.
Objective:
Quantitative computed tomography (QCT) measurements of volumetric bone mineral density (vBMD) are subject to errors due to variations in the amount of bone marrow adipose tissue (BMAT). The purpose of our study was to describe and validate a novel method to correct lumbar spine trabecular vBMD measurements for BMAT using chemical shift-encoded magnetic resonance imaging (CSE-MRI).
Methods:
CSE-MRI measurements of proton density fat fraction (PDFF) were used to correct QCT spine vBMD measurements for BMAT based on the H2O and K2HPO4 basis set equivalent densities of bone, red and yellow bone marrow. BMAT corrected and uncorrected vBMD measurements of the L1 vertebra were compared with dual-energy QCT (DEQCT) measurements in 18 subjects (mean age: 68 y, range 60 to 93 y). A further 400 subjects (mean age: 53 y, range 21 to 82 y) had 120 kVp single-energy QCT and CES-MRI scans of L2-L4 and the data used to simplify the adipose tissue correction by deriving a linear equation between the CSE-MRI vBMD correction and fractional BMAT content.
Results:
Application of the CSE-MRI derived vBMD correction changed the bias (95% limits of agreement) compared with DEQCT from 26.7 (11.0 to 42.4) mg/cm3 to 2.2 (-9.5 to 13.9) mg/cm3 at 80 kVp, and from 22.4 (3.3 to 41.6) mg/cm3 to 2.9 (-12.6 to 18.4) mg/cm3 at 120 kVp. Data for the 400 subjects gave the following relationship valid at 120 kVp: vBMD correction (mg/cm3) = -12.96 + 75.76 × BMAT.
Conclusion:
CSE-MRI measurements of PDFF can be used to correct for BMAT content and improve the accuracy of lumbar spine QCT vBMD measurements calibrated using a K2HPO4 phantom.
Related Concept Videos
Distance Corrections
Power Factor Correction
NMR Spectrometers: Resolution and Error Correction
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Bone Marrow Sampling and Transplants
The transplant begins with high doses of chemotherapy and radiation treatment, which aim to destroy...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...