Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
Maximum Swing Flexion or Gait Symmetry: A Comparative Evaluation of Control Targets on Metabolic Energy Expenditure
Wujing Cao1,2, Weiliang Zhao1,2, Hongliu Yu1,2
1Rehabilitation Engineering and Technology Institute, University of Shanghai for Science and Technology, Shanghai 200093, China.
Background:
The metabolic energy expenditure (MEE) was the most important assessment standard of intelligent prosthetic knee (IPK). Maximum swing flexion (MSF) angle and gait symmetry (GS) were two control targets representing different developing directions for IPK. However, the few comparisons based on MEE assessment between the MSF and GS limited the development of the IPK design.
Objectives:
The aim of the present work was to find out the MEE difference of amputees using IPK with control targets of MSF and GS and determine which target was more suitable for the control of IPK based on the MEE assessment.
Methods:
The crossover trial was designed. Six unilateral transfemoral amputees participated in the study. The amputees were assessed when wearing the IPK with different control targets, namely, the maximum swing flexion angle and gait symmetry. The oxygen consumption analysis during walking at different speeds on a treadmill was carried out.
Results:
All subjects showed increased oxygen consumption as walking speed increased. However, no statistically significant differences were found in oxygen consumption for different control targets. The ANOVA test showed that the overall effects of the control targets of the prosthetic knee on oxygen consumption were not significant across all walking speeds.
Conclusions:
The control targets of MSF and GS showed no significant differences on MEE in above-knee amputees using IPK. From perspective of amputee's metabolic costs, either maximum swing flexion or gait symmetry could be suitable control target for the IPK.
More Related Videos
06:58A Structured Rehabilitation Protocol for Improved Multifunctional Prosthetic Control: A Case Study
Published on: November 6, 2015
11:16Engineering Platform and Experimental Protocol for Design and Evaluation of a Neurally-controlled Powered Transfemoral Prosthesis
Published on: July 22, 2014
Related Concept Videos
The Swing Equation
In a steady-state operation, the mechanical torque (Τm) supplied to the generator is balanced by the electrical torque (Τe)...
What is Metabolism?
Symmetry
Conservation of Energy in Control Volume
For steady flow systems, the time derivative of the stored energy becomes zero since there is no energy accumulation within the control volume. This simplifies the energy equation to:
Free Energy
Maximum Deflection
The maximum deflection occurs at a specific point, known as point O, where the tangent to the deflection curve is horizontal. To find point O, the slope of the tangent at any...