In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
A Novel Toroidal-Flow Left Ventricular Assist Device Minimizes Blood Trauma: Implications of Improved Ventricular
Carlo R Bartoli1, Samson Hennessy-Strahs1, Jeff Gohean2
1Division of Cardiovascular Surgery, Department of Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania.
Background:
Continuous-flow left ventricular assist devices (LVADs) cause blood trauma that includes von Willebrand factor degradation, platelet activation, and subclinical hemolysis. Blood trauma contributes to bleeding, thrombosis, and stroke, which cause significant morbidity and mortality. The TORVAD (Windmill Cardiovascular Systems, Inc, Austin, TX) is a first-of-its kind, toroidal-flow LVAD designed to minimize blood trauma. We tested the hypothesis that the TORVAD causes less blood trauma than the HeartMate II (Abbott Laboratories, Pleasanton, CA) LVAD.
Methods:
Whole human blood was circulated for 6 hours in ex vivo circulatory loops with a HeartMate II (n = 8; 10,000 rpm, 70 ± 6 mm Hg, 4.0 ± 0.1 L/min) or TORVAD (n = 6; 144 rpm, 72 ± 0.0 mm Hg, 4.3 ± 0.0 L/min). von Willebrand factor degradation was quantified with electrophoresis and immunoblotting. Platelet activation was quantified by cluster of differentiation (CD) 41/61 enzyme-linked immunosorbent assay (ELISA). Hemolysis was quantified by plasma free hemoglobin ELISA.
Results:
The TORVAD caused significantly less degradation of high-molecular-weight von Willebrand factor multimers (-10% ± 1% vs -21% ± 1%, p < 0.0001), accumulation of low-molecular-weight von Willebrand factor multimers (22% ± 2% vs 45% ± 2%, p < 0.0001), and accumulation of von Willebrand factor degradation fragments (7% ± 1% vs 25% ± 6%, p < 0.05) than the HeartMate II. The TORVAD did not activate platelets, whereas the HeartMate II caused significant platelet activation (CD 41/61: 645 ± 20 ng/mL vs 1,581 ± 150 ng/mL, p < 0.001; normal human CD 41/61, 593 ng/mL; range, 400 to 800 ng/mL). Similarly, the TORVAD caused minimal hemolysis, whereas the HeartMate II caused significant hemolysis (plasma free hemoglobin: 11 ± 2 vs 109 ± 10 mg/dL, p < 0.0001; normal human plasma free hemoglobin <4 mg/dL).
Conclusions:
The TORVAD design, with markedly lower shear stress and pulsatile flow, caused significantly less blood trauma than the HeartMate II. LVADs with reduced blood trauma are likely to improve clinical outcomes and expand LVAD therapy into patients with less advanced heart failure.
Related Concept Videos
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...
Mnemonic Devices
Acronyms
Acronyms are created by using the initial letters of a series of words to form a new word or phrase. This approach condenses complex information into a single, memorable entity. For example,...
Toroids
When connected to a supply, the magnetic field generated in the toroid has field lines circular and concentric to its axis. Conventionally, the direction of this magnetic field is expressed using the right-hand rule. If the fingers of the right hand curl in the current direction, the thumb points in...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Blood Flow
Matrix-Assisted Laser Desorption Ionization (MALDI)