Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Underfilled Balloon-Expandable Transcatheter Aortic Valve Implantation With Ad Hoc Post-Dilation - Pulsatile Flow
Masahiro Yamawaki1,2, Kazuto Obama3, Saeko Sasuga4
1Cooperative Major in Advanced Biomedical Sciences, Joint Graduate School of Tokyo Women's Medical University and Waseda University, Waseda University.
Background:
Underfilled transcatheter aortic-valve implantation with ad hoc post-dilation is a therapeutic option for patients with borderline annuli to avoid acute complication. The effects of this technique on valve leaflet behavior, hydrodynamic performances, and paravalvular leakage (PVL) using patient-specific three-dimensional (3D) aortic-valve models were investigated.
Methods And Results:
A female octogenarian patient was treated with this technique by using a 23-mm Sapien-XT. Patient-specific models were constructed from pre-procedure computed tomography (CT) data. Change in aortic annulus areas during systolic/diastolic phases and post-procedure stent areas were adjusted to those of the patient. The following was performed: (1) -3 cc initial and -2 cc underfilled post-dilation to the scale-down model by adjusting percent oversizing; and (2) -1 cc initial underfilling, nominal volume, and repeat nominal volume post-dilation using the patient-specific model. Underfilling was associated with higher %PVL. Observation using a high-speed camera revealed distorted leaflets after underfilled implantation, with a longer valve-closing time and smaller effective orifice areas, especially in the -3 cc underfilled implantation. Micro-CT analysis revealed that the transcatheter valves shifted to the opposite side of the large annulus calcification after post-dilation and reduced the malapposition there.
Conclusions:
Excessive underfilled implantation showed unacceptable acute hemodynamics. Abnormal leaflet motions after underfilled implantation raised concerns about durability. Flow simulations using patient-oriented 3D models could help to investigate hemodynamics, leaflet motions, and the PVL mechanism.
Related Concept Videos
Heat Flow and Specific Heat
Cardiomyopathy II: Dilated Cardiomyopathy
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Mitral Valve Prolapse I: Introduction
Molecular Models