Automated oxygen control with O2matic® during admission with exacerbation of COPD
Ejvind Frausing Hansen1, Jens Dahlgaard Hove1, Charlotte Sandau Bech1
1Medical Unit, Amager and Hvidovre Hospital, Copenhagen, Denmark, ejvind.frausing.hansen@regionh.dk.
Purpose:
It is a challenge to control oxygen saturation (SpO2) in patients with exacerbations of COPD during admission. We tested a newly developed closed-loop system, O2matic®, and its ability to keep SpO2 within a specified interval compared with manual control by nursing staff.
Patients And Methods:
We conducted a crossover trial with patients admitted with an exacerbation of COPD and hypoxemia (SpO2 ≤88% on room air). Patients were monitored with continuous measurement of SpO2. In random order, they had 4 hours with manually controlled oxygen and 4 hours with oxygen delivery controlled by O2matic. Primary outcome was time within a prespecified SpO2 target interval. Secondary outcomes were time with SpO2 <85%, time with SpO2 below target but not <85%, and time with SpO2 above target.
Results:
Twenty patients were randomized and 19 completed the study. Mean age was 72.4 years and mean FEV1 was 0.72 L (33% of predicted). Patients with O2matic-controlled treatment were within the SpO2 target interval in 85.1% of the time vs 46.6% with manually controlled treatment (P<0.001). Time with SpO2 <85% was 1.3% with O2matic and 17.9% with manual control (P=0.01). Time with SpO2 below target but not <85% was 9.0% with O2matic and 25.0% with manual control (P=0.002). Time with SpO2 above target was not significantly different between treatments (4.6% vs 10.5%, P=0.2). Patients expressed high confidence and a sense of safety with automatic oxygen delivery.
Conclusion:
O2matic was able to effectively control SpO2 for patients admitted with an exacerbation of COPD. O2matic was significantly better than manual control to maintain SpO2 within target interval and to reduce time with unintended hypoxemia.
Related Concept Videos
COPD: Pathogenesis and Clinical Features
The primary cause for the onset of COPD is cigarette smoking and exposure to air pollution. These hazardous factors initiate a chain reaction within the lungs, resulting in chronic inflammation, damage to the airways, and a...
COPD: Management Using Bronchodilators and Corticosteroids
Oxygen Transport in the Blood
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygenic Photosynthesis
Oxygen Requirements and Growth Patterns