Comparison of a micro-electro-mechanical system airflow sensor with the pneumotach in the forced oscillation
Xiaohe K Xu1, Brian P Harvey2, Kenneth R Lutchen2
1Feather Sensors, LLC, Millville, NJ 08332, USA, ned@feathersensor.com.
Purpose:
This study supports the use of thin-film micro-electro-mechanical system (MEMS) airflow sensors in the forced oscillation technique.
Materials And Methods:
The study employed static testing using air flow standards and computer-controlled sound attenuations at 8 Hz. Human feasibility studies were conducted with a testing apparatus consisting of a pneumotach and thin-film MEMS air flow sensors in series. Short-time Fourier transform spectra were obtained using SIGVIEW software.
Results:
Three tests were performed, and excellent correlations were observed between the probes. The thin-film MEMS probe showed superior sensitivity to higher frequencies up to 200 Hz.
Conclusion:
The results suggest that lower-cost thin-film MEMS can be used for forced oscillation technique applications (including home care devices) that will benefit patients suffering from pulmonary diseases such as asthma, COPD, and cystic fibrosis.
Related Concept Videos
Electro-mechanical Systems
A key component of the DC motor is the armature, a rotating circuit positioned within a magnetic field. As an electric current passes through the...
Forced Oscillations
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Cell-matrix's Response to Mechanical Forces
Anchoring junctions mechanically attach a cell to the...
Intermolecular Forces
Oscillations In An LC Circuit