Measurement of knee joint range of motion with a digital goniometer: A reliability study
Melanie Svensson1,2, Veronika Lind1, Marita Löfgren Harringe1,2
1Hela Kroppen Fysioterapi och Friskvård AB, Stockholm, Sweden.
Objectives:
Measurements of joint range of motion (ROM) are part of a physical therapist's daily work. Activities of daily living and exercises can be complicated to perform when ROM is limited, and depending on the demands in daily living, the knee joint requires different ROM. In sports, a few degrees in ROM may make the difference between getting injured or not. The goals for physical therapists are to help the patients to regain full ROM, mobility, strength, and function after sustaining an injury. To measure joints with the manual universal goniometer is considered time-consuming and difficult with respect to repeated measurements. Recently, a new digital instrument for measuring ROM was developed-EasyAngle. A first objective of the study was to investigate the reliability of EA for measuring knee joint angles, considering intrarater and interrater reliability. A second objective was to investigate if there were any differences in the intrarater reliability between a novice and an experienced assessor.
Method:
Passive knee angles were measured in fixed positions for 40 knee joints (20 subjects). Two registered physical therapists, one novice and one experienced, conducted the measurements. Both registered physical therapists were blinded to the measurements throughout the study.
Results:
The results showed very good intrarater (intraclass correlation coefficient [ICC] 0,997-0,998, standard error of mean 1.15-1.48, smallest detectable difference [SDD] 3.19-4.09, limits of agreement -3.36-3.04, -4.66-4.09) and interrater reliability (ICC 0.994, standard error of mean 2.11, SDD 5.85, limits of agreement -4.75-6.95) for measurements of knee joint ROM. No statistical difference between a novice and an experienced assessor was detected (p = 0.86).
Conclusion:
The results of the present study showed very good ICC values for both intrarater and interrater reliability measuring knee joint ROM with EasyAngle. Relatively high SDD values were seen for both assessors and may indicate a problem monitoring small differences between measurements. Further studies are recommended to increase the generalizability of the results.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Reliability and Validity
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Variation: Normal Distribution, Range, and Standard Deviation
Structural Joints: Synovial Joints
Structural Joints: Fibrous Joints
Suture
All the bones of the skull, except for the mandible, are joined to each other by a fibrous joint called a suture. The fibrous connective tissue found at a suture strongly unites the adjacent skull bones and thus helps to protect the brain and form the face. In...