Comparison of axial length using a new swept-source optical coherence tomography-based biometer - ARGOS with partial
Jameel Rizwana Hussaindeen1, Ephrame G Mariam2, Sushil Arunachalam2
1Myopia Clinic, Srimathi Sundari Subramanian Department of Visual Psychophysics, Unit of Medical Research Foundation, Sankara Nethralaya, Nungambakkam, Chennai, India.
Purpose:
To compare the axial length measurements obtained by a new swept source optical coherence tomography based biometer-ARGOS with partial coherence interferometry based biometer -IOL master in school children between the ages of 11-17.
Methods:
A prospective, cross-sectional, device comparison study was conducted in a school vision screening program comparing the axial length (AL) and corneal curvature (K) measurements obtained by two biometers- ARGOS and IOL master. Children with 6/9 vision or better, without any ocular abnormalities were included in the study. Two trained optometrists performed the measurements and were masked for the outcome measures.
Results:
The sample size was 188 with a mean(SD) age of 13.88±1.69 years, of which 101 were boys. The mean (SD) AL was 23.94± 1.01 mm with Argos and 23.83 ± 1.03 mm with IOL Master (paired t-test, p>0.05). The mean K was 43.62D±1.59 with Argos and 43.64D±1.61 with IOL master (paired t-test, p>0.05). There was a strong positive correlation between the biometers for AL (r = 1.00, p<0.0001) and K (r = 0.99, p<0.0001). The mean difference in axial length between the two biometers was 0.11± 0.04 mm and the limits of agreement were between -0.02 to -0.19. The mean difference in corneal curvature was 0.02±0.15D and the limits of agreement were between -0.28 to 0.32D.
Conclusion:
Axial length measurements using SS-OCT and PCI based biometers were in agreement and comparable among children between the ages of 11 to 17.
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
Weak Base Solutions
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Bronsted-Lowry Acids and Bases
Axial and Appendicular Muscles
Axial Muscles
Axial muscles, situated along the body's midline, are intricately connected to the axial skeleton, which includes the skull, spine, ribs, and sternum. These muscles facilitate facial expressions and...
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...