EFFECTIVE DOSE ESTIMATION FROM ORGAN DOSE MEASUREMENTS IN FAST-kV SWITCH DUAL ENERGY COMPUTED TOMOGRAPHY
Justin Raudabaugh1,2, Giao Nguyen2, Carolyn Lowry3
1Medical Physics Graduate Program, Duke University, Durham, NC, USA.
Abstract:
The purpose of this study was to validate a novel approach to estimating effective dose (E) in 'fast-kV switch dual energy computed tomography' using MOSFET detectors. The effective energy of the combined dual energy environment was characterized with the dual energy CT scanner and then MOSFETs were calibrated matching to the effective energy of the dual energy CT beam with a conventional CT beam. The calibration method was then experimentally validated by comparing the dose between MOSFET and an ion chamber (IC) using a standard CTDI body phantom. The measured doses of the MOSFET and IC were 17.1 mGy ± 3.8% and 17.1 mGy ± 0.4%, respectively. To measure organ doses, an adult anthropomorphic phantom loaded with 18 MOSFET detectors was scanned using a standard fast-kV switch dual energy abdomen/pelvis CT protocol. E was calculated by applying ICRP 103 tissue weighting factors as well as partial volume correction factors for organs that were not completely covered by the protocol field-of-view. E from the dual energy abdomen/pelvis CT was calculated to be 17.8 mSv ± 11.6%. This calculation was then compared to E from dose length product method, which yielded 14.62 mSv.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dose-Response Relationship: Overview