Live human assessment of depth-dependent corneal displacements with swept-source optical coherence elastography
Vinicius S De Stefano1,2, Matthew R Ford1, Ibrahim Seven1
1Cole Eye Institute, Cleveland Clinic, Cleveland, OH, United States of America.
Purpose:
To assess depth-dependent corneal displacements in live normal subjects using optical coherence elastography (OCE).
Methods:
A corneal elastography method based on swept-source optical coherence tomography (OCT) was implemented in a clinical prototype. Low amplitude corneal deformation was produced during OCT imaging with a linear actuator-driven lens coupled to force transducers. A cross-correlation algorithm was applied to track frame-by-frame speckle displacement across horizontal meridian scans. Intra-measurement force and displacement data series were plotted against each other to produce local axial stiffness approximations, k, defined by the slope of a linear fit to the force/displacement data (ignoring non-axial contributions from corneal bending). Elastographic maps displaying local k values across the cornea were generated, and the ratio of mean axial stiffness approximations for adjacent anterior and posterior stromal regions, ka/kp, was calculated. Intraclass correlation coefficients (ICC) were used to estimate repeatability.
Results:
Seventeen eyes (ten subjects) were included in this prospective first-in-humans translational study. The ICC was 0.84. Graphs of force vs. displacement demonstrated that, for simultaneously acquired measurements involving the same applied force, anterior stromal displacements were lower (suggesting stiffer behavior) than posterior stromal displacements. Mean ka was 0.016±0.004 g/mm and mean kp was 0.014±0.004 g/mm, giving a mean ka/kp ratio of 1.123±0.062.
Conclusion:
OCE is a clinically feasible, non-invasive corneal biomechanical characterization method capable of resolving depth-dependent differences in corneal deformation behavior. The anterior stroma demonstrated responses consistent with stiffer properties in compression than the posterior stroma, but to a degree that varied across normal eyes. The clinical capability to measure these differences has implications for assessing the biomechanical impact of corneal refractive surgeries and for ectasia risk screening applications.
More Related Videos
07:51Full-Field Optical Coherence Microscopy for Histology-Like Analysis of Stromal Features in Corneal Grafts
Published on: October 21, 2022
03:47Retinal and Choroidal Thickness Changes in Populations with Helicobacter pylori Infection by Swept-Source Optical Coherence Tomography
Published on: November 1, 2024
Related Concept Videos
Independent and Dependent Sources
Independent voltage or current sources supply a fixed amount of voltage or current, respectively, which is unaffected by other elements within the circuit. These are represented using specific symbols. Independent voltage sources are symbolized with polarities (+ and -), indicating the direction of the...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Displacement Current
Position and Displacement
Position and Displacement Vectors
Further, several important kinds of...
Significance of Displacement Current