Related Experiment Video
Updated: Jan 31, 2026

Intra-Operative Behavioral Tasks in Awake Humans Undergoing Deep Brain Stimulation Surgery
Published on: January 6, 2011
Implementation of Intraoperative Computed Tomography for Deep Brain Stimulation: Pitfalls and Optimization of
Barbara Carl1, Miriam Bopp2, Marko Gjorgjevski1
1Department of Neurosurgery, University Marburg, Marburg, Germany.
Objective:
Deep brain stimulation (DBS) is an effective treatment for movement disorders. Stereotactic electrode placement can be guided by intraoperative imaging, which also allows for immediate intraoperative quality control. This article is about implementation and refining a workflow applying intraoperative computed tomography (iCT) for DBS.
Methods:
Eighteen patients underwent DBS with bilateral implantation of directional electrodes applying a 32-slice movable computed tomography scanner in combination with microelectrode recording.
Results:
iCT led to a significant decrease in overall procedural time, despite performing multiple scans. In 3 of the initial 5 cases, iCT caused an adjustment of the final electrodes demonstrating the learning curve and the necessity to integrate road mapping for the exchange of microelectrode to final electrode. Implementation of low-dose computed tomography protocols added microelectrode iCT to the refined workflow, resulting in an intraoperative adjustment of a trajectory in 1 patient. Low-dose protocols lowered the total effective dose to 1.15 mSv, that is, a reduction by a factor of 3.5 compared to a standard non-iCT DBS procedure, despite repeated iCTs. Intraoperative lead detection based on final iCT revealed a radial error of 1.04 ± 0.58 mm and a vector error of 2.28 ± 0.97 mm compared to the preoperative planning, adjusted by the findings of microelectrode recording.
Conclusions:
iCT can be easily integrated into the surgical workflow resulting in an overall efficient time-saving procedure. Repeated intraoperative scanning ensures reliable electrode placement, although low-dose scanning protocols prevent extensive radiation exposure. iCT of microelectrodes is feasible and led to the adjustment of 1 electrode.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Biological Effects of Radiation
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Nursing Implementation
The five steps to implementing effective nursing care include reassessing the patient, reviewing and revising the existing nursing care plan, organizing the resources and care delivery, anticipating and preventing complications, and implementing nursing interventions.