UPLC-MS/MS method for the simultaneous determination of imatinib, voriconazole and their metabolites concentrations
Ren-Ai Xu1, Qianmeng Lin2, Xiangjun Qiu3
1The First Affiliated Hospital of Wenzhou Medical University, 325000, Wenzhou, PR China.
Abstract:
In the present study, a simple ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method used to measure the plasma concentrations of imatinib, voriconazole and their metabolites (N-desmethyl imatinib and N-oxide voriconazole) in rats simultaneously making use of diazepam as the internal standard (IS) had been developed and validated. A simple protein precipitation by acetonitrile was employed for the sample preparation, then the analytes (imatinib, voriconazole and their metabolites) were eluted on an Acquity UPLC BEH C18 column (2.1 mm × 50 mm, 1.7 μm) using the mobile phase that made up by acetonitrile (A) and 0.1% formic acid in water (B). In positive ion mode, four analytes and IS were monitored by multiple reaction monitoring (MRM) as the following mass transition pairs: m/z 494.3→394.2 for imatinib, m/z 480.3→394.2 for N-desmethyl imatinib, m/z 350.1→281.1 for voriconazole, m/z 366.1→224.1 for N-oxide voriconazole, and m/z 285.0→154.0 for IS. This method exhibited a good linearity for each analyte. Inter-day and intra-day precision were determined with values of 0.3-14.8% and 2.6-14.8%, respectively; the accuracy values were from -12.5% to 10.2%. Finally, data of matrix effect, extraction recovery, and stability were all conformed to the bioanalytical method validation of acceptance criteria of FDA recommendations. This method is an efficient tool for simultaneous determination of the four analytes and has been successfully applied for pharmacokinetic study in rats.
More Related Videos
11:17A Simple Fractionated Extraction Method for the Comprehensive Analysis of Metabolites, Lipids, and Proteins from a Single Sample
Published on: June 1, 2017
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Related Concept Videos
Concentration Cells
Consider the following voltaic cell:
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Protein-Drug Binding: Determination Methods
Indirect methods involve isolating the bound drug from its free form in biological samples such as blood, serum, or plasma. These techniques aim to measure the percentage of drugs bound to proteins. Equilibrium dialysis is a commonly used method where the free drug concentration at equilibrium is measured by separating the bound...
Calculating Equilibrium Concentrations
A more...
Concentration and Rate Law
For example, in a generic reaction aA + bB ⟶ products, where a and b are stoichiometric coefficients, the rate law can be written as:
Solution Concentration and Dilution