Continuous Direct Left Atrial Pressure: Intraprocedural Measurement Predicts Clinical Response Following MitraClip
Shingo Kuwata1, Maurizio Taramasso1, Albert Czopak1
1University Heart Center Zürich, University Hospital of Zurich, Zurich, Switzerland.
Objectives:
The aim of this study was to assess the additional utility of measuring left atrial (LA) pressure as a complement to transesophageal echocardiography (TEE) and Doppler imaging.
Background:
The efficacy of the MitraClip (MC) is assessed intraoperatively by TEE. However, TEE measures are operator dependent and can be influenced by left ventricular (LV) function and changes in LA compliance.
Methods:
Fifty patients undergoing MC therapy with continuous left-sided heart pressure measurements were analyzed. LA V-wave pressure (LAvP), LA mean pressure (LAmP), LV systolic pressure, and LV end-diastolic pressure were measured continuously. LA pressures were indexed to LV pressures to account for changes in afterload during the procedure.
Results:
Most patients (70%) had degenerative mitral regurgitation (MR). TEE MR grade decreased from 3+ to 0+. LAvP (p < 0.001), LAmP (p = 0.007), LV end-diastolic pressure (p = 0.001), LAvP index (p < 0.001), and LAmP index (p = 0.001) decreased significantly, and LV systolic pressure(p = 0.009) significantly increased after MC therapy. In multivariate Cox regression analysis, intraprocedural increase of LAmP index, but not post-MC ≥2+ residual MR, was significantly associated with rehospitalization due to heart failure (hazard ratio: 3.377; 95% CI: 3.180 to 3.585; p = 0.007) and with New York Heart Association functional class III to IV (hazard ratio: 1.497; 95% CI: 1.006 to 2.102; p = 0.005) in the follow-up period.
Conclusions:
This study demonstrates the value of real-time monitoring of LA pressure during MC therapy to predict clinical outcomes. An increase in LAmP was a predictive of worse clinical outcomes at short-term follow-up, independent from echocardiographic findings.
More Related Videos
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Continuous Renal Replacement Therapy
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...