Related Experiment Video
Updated: Jan 31, 2026

Author Spotlight: Innovative Methodology for Implanting and Securing Neural Probes in the Rodent Spinal Cord
Published on: July 12, 2024
Device implantation for patients on antiplatelets and anticoagulants: Use of suction drain
Sanjeev S Mukherjee1, Daljeet Saggu1, Sridevi Chennapragada1
1Department of Cardiology, CARE Hospital, Banjara Hills, Hyderabad, India.
Background And Objectives:
Cardiovascular implantable electronic devices (CIED) are frequently implanted in patients on anti-thrombotic agents. Pocket hematomas are more likely to occur in these patients. The use of a sterile surgical drain in the pulse generator pocket site could prevent hematomas, but fear of infection precludes its use. The objective of the present study is to study the safety and efficacy of surgical drain in patients on antithrombotics undergoing CIED implantations.
Methods:
This is a single-centre, retrospective study involving patients undergoing CIED implantations on antithrombotics (antiplatelets and anticoagulants) from August 2013 to July 2016. Patients with high risk of thromboembolism were continued on oral antithrombotics or were bridged with heparin after stopping oral antithrombotics. A sterile close wound suction drain was placed in device pockets following CIED implantations. Post procedure, pressure dressing was applied and removed after 12 h once the drain volume was less than 10 ml in 24 h.
Results:
Sixty seven patients required surgical drain implantation. Major indications for antithrombotic use were presence of intracoronary stent, atrial fibrillation and mechanical valve replacements. The mean post-procedural hospital stay was 3 ± 0.9 days and mean overall drain was 16.6 ± 8.2 ml. At a mean follow up of 17.6 ± 8.2 months, one patient (1.4%) had pocket hematoma. There were no infections.
Conclusion:
The use of a surgical drain in CIED implantation significantly reduces the risk of hematoma formation without increasing the risk of infection. Antithrombotic drugs can be safely continued at the time of implantation of cardiac devices.
Related Concept Videos
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Tracheostomy Suctioning II: Procedure
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...