Kinematic alignment technique for medial OXFORD UKA: An in-silico study
Charles Rivière1, Ciara Harman2, Anthony Leong3
1The MSK Lab-Imperial college London, South West London Elective Orthopaedic Centre, London, United Kingdom.
Background:
Mobile bearing unicompartmental knee arthroplasty (UKA) Oxford™ components are recommended to be systematically and mechanically aligned (MA) for restoring the constitutional lower-limb alignment. Good long-term clinical outcomes have been generated with the medially implanted MA Oxford™, but some sub-optimal biomechanical-related complications still remain. Kinematic Alignment (KA) is a personalised technique for anatomically and kinematically implanting components (total knee, fixed bearing partial knee, total hip) aimed at creating more physiological prosthetic joint biomechanics. Interestingly, for decades the principles for implanting fixed bearing UKA components were consistent with those promoted by the KA technique, but differently formulated. We initiated this computational study to assess the feasibility of this technique with the Oxford™ components, as we thought this more anatomical implantation may be clinically advantageous.
Hypothesis:
We surmised that kinematically aligning the Oxford™ medial UKA would maximise the prosthesis-bone interface through maximising the implants' size used (question 1), and alter, within an acceptable limit, the components' orientation (question 2) compared to conventional mechanical alignment.
Methods:
A cohort of 40 consecutive medial osteoarthritic knee patients scheduled for UKA had a preoperative CT scan that was segmented to create 3D knee bone models. MA and KA of medial UKA Oxford® components (Zimmer-Biomet, Warsaw, Indiana, USA) were simulated. Component sizing and positioning were compared between the two techniques.
Results:
We found no difference in component size, but significantly fewer occurrences of borderline fit with the KA simulation. KA technique oriented the femoral component 3.6° more valgus (from 1° varus to 7° valgus) and the tibial component 2.9° more varus (from 8° varus to 0°) compared to the MA technique. The tibial component slope in KA simulation was 6.4° posterior (from 0 to 12°) compared to a systematic 7° posterior for MA positioning.
Discussion And Conclusion:
Kinematic alignment of the medial Oxford™ generated a different, albeit still acceptable (Oxford group recommendations), implant orientation, in addition to a likely better shape-fit between components and the supportive bone cut, compared to the MA technique. The potential to improve the implants' interaction and to restore a more physiological bone loading makes the KA of Oxford™ an attractive, potentially clinically beneficial option. Clinical investigations are needed to assess its true value.
Level Of Evidence:
I, computational study.
More Related Videos
Related Concept Videos
Impression Management Techniques III: Aligning Actions
Kinematic Equations - I
Kinematic Equations - II
Suppose a car merges into freeway traffic on a 200 m long ramp. If its initial velocity is 10 m/s and it accelerates at 2 m/s2, then the...
Kinematic Equations for Rotation
For instance, imagine a point A on a rigid body engaged in circular motion. The translational velocity of this particular point can be calculated by taking the time derivatives of the displacement equation, which essentially measures the...
Kinematic Equations - III
Using the kinematic equations,...
Kinematic Equations: Problem Solving