Peak sharpening limits of solvent-assisted post-column refocusing to enhance detection limits in liquid
Vincent Pepermans1, Jelle De Vos1, Sebastiaan Eeltink1
1Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, B-1050, Brussels, Belgium.
Abstract:
We report on a numerical and experimental study of the limits of peak refocusing and concentration enhancement that can be obtained with solvent front-assisted peak remobilization in a trap column receiving peaks eluting from a preceding analytical column. It is shown that the upper limit of peak refocusing can best be pursued by injecting a sufficiently large volume in a sufficiently narrow capillary and elute it with a sufficiently steep (ballistic) gradient. Corresponding equations offering a quantitative description have been derived and verified experimentally. For the latter purpose, peak volumes of the order of 0.5 to 2.0 μL were refocused in a dedicated set-up capable of trapping μL-sized peaks in a 75 μm i.d. capillary and remobilized using a nano-LC pump propelling an acetonitrile/isopropyl alcohol mixture with a viscosity matching that of the trapping solvent. Injecting 2.0 μL peaks, a peak refocusing factor of 17.3 could be achieved.
More Related Videos
10:21Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
09:45Author Spotlight: Detecting Low-Abundant Host Cell Proteins in Drug Products Using Enrichment Beads and Limited Digestion
Published on: January 19, 2024
Related Concept Videos
Limiting Reactant
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
The Number e as a Limit
Types of Limits I
Limit Laws I
Introduction to Limits