Comparison of patient dose from routine multi-phase and dynamic liver perfusion CT studies taking into account the
Kostas Perisinakis1, Antonis Tzedakis2, Styliani Pouli3
1Department of Medical Physics, Medical School, University of Crete, P.O. Box 2208, Heraklion 71003, Crete, Greece; Department of Medical Physics, University Hospital of Heraklion, P.O. Box 1352, Heraklion 71110, Crete, Greece.
Objectives:
To accurately determine and compare patient radiation burden from routine multi-phase CT (MPCT) and dynamic CT liver perfusion (CTLP) studies taking into account the effect of iodine uptake of exposed tissues/organs.
Materials And Methods:
40 consecutive MPCT of upper abdomen and 40 consecutive CTLP studies performed on a modern CT scanner were retrospectively studied. Iodine uptake of radiosensitive tissues at the time of acquisition was calculated through the difference of tissues' CT numbers between NECT and CECT images. Monte Carlo simulation and mathematical anthropomorphic phantoms were employed to derive patient-size-specific organ dose data from each scan involved taking into account the effect of iodinated contrast uptake on absorbed dose. Effective dose estimates were derived for routine multiphase CT and CTLP by summing up the contribution of NECT and CECT scans involved.
Results:
The mean underestimation error in organ doses from CECT exposures if iodine uptake is not encountered was found to be 2.2%-38.9%. The effective dose to an average-size patient from routine 3-phase CT, 4-phase CT and CTLP studies was found to be 20.6, 27.7 and 25.8 mSv, respectively. Effective dose from CTLP was found lower than 4-phase CT of upper abdomen irrespective of patient body size. Compared to 3-phase CT, the radiation burden from CTLP was found to be higher for average size-patients but again lower for overweight patients.
Conclusions:
Modern CT technology allows CTLP studies at comparable or even lower patient radiation burden compared to routine multi-phase liver CT imaging.
More Related Videos
09:02Percutaneous Hepatic Perfusion PHP with Melphalan as a Treatment for Unresectable Metastases Confined to the Liver
Published on: July 31, 2016
09:29Technique of Subnormothermic Ex Vivo Liver Perfusion for the Storage, Assessment, and Repair of Marginal Liver Grafts
Published on: August 13, 2014
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Drug Administration and Therapy Phases: Overview
The pharmaceutical phase focuses on leveraging the physicochemical properties of the drug to design and manufacture an effective product. Variants include orally administered tablets or capsules, topical creams or ointments, and parenteral-delivery solutions or emulsions.
The pharmacokinetic phase...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT