Assessment of Postoperative Morphologic Retinal Changes by Optical Coherence Tomography in Recipients of an
Stanislao Rizzo1, Laura Cinelli1, Lucia Finocchio1
1Azienda Ospedaliera Universitaria Careggi, Dipartimento di Medicina e Chirurgia Traslazionale, University of Florence, Florence, Italy.
Importance:
The postoperative retinal changes at the interface between an implant electrode array and the retina and whether these anatomic changes have an association with the patient visual performance are unknown.
Objective:
To report morphologic changes in recipients of an Argus II Retinal Prosthesis.
Design, Setting, And Participants:
This consecutive, noncomparative case series study included a retrospective review of the preoperative and postoperative optical coherence tomography of 33 eyes among 33 individuals who underwent Argus II Retinal Prosthesis System implantation between October 28, 2011, and June 8, 2017, at 2 different centers, by the same surgeon (S.R.). Thirteen patients received an implant at Azienda Ospedaliero Universitaria Pisana, Pisa, Italy, between October 28, 2011, and October 27, 2014, and 20 patients underwent surgery at Azienda Ospedaliera Universitaria Careggi, Florence, Italy, between December 20, 2014, and June 8, 2017. Patients were excluded if they did not reach the 6-month follow-up.
Main Outcomes And Measures:
All patients were evaluated before surgery, during the first postoperative day, and at 1, 3, 6, 12, and 24 months (subsequently once a year, except for patient-related adverse events), with a comprehensive ophthalmic examination, retinal fundus photography, spectral-domain optical coherence tomography, and visual function tests to evaluate the stability or improvement of their visual performance.
Results:
Of the 20 patients included in the analysis, all were of white race/ethnicity, 12 (60%) were male, and the mean (SD) age was 57.4 (11.6) years. Optical coherence tomography revealed the development of a fibrosislike hyperreflective tissue limited at the interface between the array and retina in 10 eyes (50%). In 9 of 10 patients (90%), fibrosis evolved and progressed to retinal schisis. Despite the development of the fibrosis and schisis, there was no deterioration in the patient's visual performance evaluated prospectively with visual function tests (square localization and direction of motion).
Conclusions And Relevance:
Optical coherence tomography may be used to observe the retinal anatomic changes in patients with an Argus II Prothesis. This analysis revealed the development of a fibrosislike hyperreflective tissue limited at the interface between array and retina that progressed to retinal schisis but with no deterioration in the patients' visual performance.
More Related Videos
10:00Techniques for Processing Eyes Implanted with a Retinal Prosthesis for Localized Histopathological Analysis: Part 2 Epiretinal Implants with Retinal Tacks
Published on: February 14, 2015
08:17Optical Coherence Tomography: Imaging Mouse Retinal Ganglion Cells In Vivo
Published on: September 22, 2017
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Electron Affinity
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Electron Transport Chains
The ETC is comprised of...
Microbial Morphologies