Resorbable Magnesium Scaffold in Coronary Bifurcations - Report of in Vitro Experiments
Gabor G Toth1, Michael Haude2, Daniel Lootz3
1University Heart Centre Graz, Medical University of Graz, Austria; Cardiovascular Research Centre Aalst, OLV-Clinic Aalst, Belgium.
Objectives:
The aim of this work is to evaluate in an in vitro setting the behavior of resorbable magnesium scaffolds (RMS) in bifurcations.
Background:
As coronary bifurcations represent an important portion of all PCIs, it is crucial to understand whether RMS is applicable in these complex lesions.
Methods:
Performance of RMS was evaluated with focus on bifurcations. In bifurcations RMS was tested for (1) main branch stenting with side branch opening and proximal optimization; for (2) main branch stenting with final kissing and proximal optimization; for (3) T-and-protrusion technique; for (4) string technique, which is a minimalistic version of conventional culotte technique. All tests were performed using 3.50 × 25 mm RMS. Results were evaluated by fluoroscopy, optical coherence tomography (OCT) and micro-computed tomography (μCT), for scaffold conformability, strut apposition, structural deformation and strut fracture.
Results:
All planned procedural steps were performed successfully with good result according to fluoroscopy. By OCT the overall malapposition rate in bifurcation cases was 4.3%, occurring predominantly in the carinal area. No malapposition was seen at the proximal main branch confirming proper conformability of RMS. μCT analysis has shown that final kissing dilation resulted in fully stretched struts in cases, where performed with 3.5 and 3.0 mm balloons. In one case a broken connector (T-and-protrusion) and in another case, a broken strut (String technique) were identified.
Conclusions:
RMS can structurally cope with bifurcations. Still, for cases and techniques where overexpansion beyond the recommended limit is needed, RMS might not be the proper device due to risk of strut fracture.
Related Concept Videos
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Data Reporting and Recording
What is an Experiment?
Types of Reports I: Hands-off Report
Following are the key components and categories of hand-off reports:
Purpose and Process:
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...