Avoiding a Lemon: Performance Consistency of Single-Use Ureteroscopes
Brenton Winship1, Daniel Wollin2, Evan Carlos1
11 Division of Urologic Surgery, Duke University Medical Center, Durham, North Carolina.
Introduction And Objectives:
Single-use flexible ureteroscopes are increasingly popular because of high repair costs of reusable ureteroscopes. As new single-use ureteroscopes enter the market, the consistency of performance of these devices has been questioned. Our objective was to compare two single-use ureteroscopes: the Pusen PU3022a (Zhuhai Pusen) and the LithoVue (Boston Scientific) with emphasis on physical and optical performance consistency.
Methods:
Ten LithoVue and 10 Pusen ureteroscopes were evaluated in never-used condition. The following parameters were recorded: maximal tip deflection with an empty working channel, 200 μm laser fiber, and 1.9F basket; image resolution at 10, 20, and 50 mm; and irrigation flow rate; all ureteroscopes were then fully deflected for 200 cycles. Maximum deflection was remeasured. Mean values for each parameter were compared. To examine within-manufacturer consistency, the variance was calculated and an F-test performed to evaluate for equivalence.
Results:
Both ureteroscopes provided max deflection over 270°. The Pusen flexed to a greater degree than LithoVue. The Pusen lost more deflection with a laser fiber in the working channel. LithoVue had higher resolution at 10 mm, but the ureteroscopes were similar at 20 and 50 mm. Although the working channel diameter is identical, irrigation flow was higher with the Pusen. However, its working channel was 7 cm shorter than LithoVue. After 200 cycles of deflections, LithoVue had less deflection loss, although one ureteroscope was excluded from analysis because of mechanical failure. Variability was minimal and not significant between manufacturers at all measured parameters.
Conclusions:
LithoVue and the newer PU3022a have similar "out-of-the-box" performance characteristics and seem durable. These two single-use flexible ureteroscopes performed consistently regardless of the manufacturer.
More Related Videos
Related Concept Videos
Frustration and Conflict: Avoidance-Avoidance, Double-Approach Avoidance
Personality Disorders: Narcissistic and Avoidant
Characteristics of Narcissistic Personality Disorder
Narcissistic individuals exhibit an inflated sense of self-importance and an excessive need for admiration. They are often...
Avoidance Learning and Learned Helplessness
Avoidance learning occurs when an organism learns that a specific behavior can prevent an unpleasant outcome. For example, a student who receives a bad grade may start studying harder to avoid future poor grades. This behavior persists even when the negative outcome is no longer present. Avoidance learning is powerful because it maintains behavior in the absence of the...
Primary Motives: Sleep, Sex, and Pain Avoidance
Sleep is a fundamental physiological drive that fosters a state of restfulness crucial for several bodily functions. It facilitates body restoration, the process by which the body repairs, rejuvenates, and maintains itself during sleep, including memory...
Frustration and Conflict: Approach-Approach, Approach-Avoidance
One common type of conflict is the Approach–Approach Conflict. In this case, a person faces two desirable...
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...