Microcirculation evaluation of facial nerve palsy using laser speckle contrast imaging: a prospective study
Yayou Zhang1, Liang Zhao2, Jiandong Li2
1Otorhinolaryngological Department, Beijing Electric Power Hospital, 100073, Beijing, China. yoyochangzyy@sina.com.
Objective:
Facial nerve palsy (FNP) is a common disease in the otorhinolaryngological department. Besides the main symptom of motionlessness occurring in the ipsilateral facial muscles in FNP, there are other common complaints of numbness, stiffness and tightness in ipsilateral face described by the patients. Based on our pilot study, we further investigated the relevance between these complaints and facial microcirculation.
Method:
Function of facial microcirculation was evaluated by laser speckle contrast imaging (LCSI). Facial perfusion was measured in 143 patients with facial nerve palsy (FNP) at the first visit and follow-up visit under the same conditions.
Results:
Difference in FNP patients' facial microvascular perfusions between ipsilateral and contralateral side was significant (P = 0.0002613). Facial perfusion of patients with Bell's palsy (P = 00089) and facial nerve tumors (P = 0.025110) was significantly decreasing in the ipsilateral side. Improvement of perfusion could be seen after treatment.
Conclusion:
A positive correlation of FNP severity and microvascular impairment can be noticed. During treatment, patients' ipsilateral perfusion could increase. Therefore, this objective method can measure ipsilateral perfusion in the patients with FNP and the ipsilateral microvascular impairment can be detected through this method.
Related Concept Videos
Muscles for Facial Expressions
Facial Feedback Hypothesis
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies VII: Vascular Imaging
Imaging Studies II: Ultrasonography
Imaging Studies I: Kidney, Ureter, and Bladder Studies